Hathout Eba, Patel Nina, Southern Christina, Hill Julie, Anderson Reginald, Sharkey Jeannine, Hadley-Scofield Merrilee, Tran Long, Leptien Amy, Lopatin Margarita, Wang Betty, Mace John, Eastman Richard
Pediatric Diabetes Center, Loma Linda University, 11175 Campus St, CP A1120R, Loma Linda, CA 92354, USA.
Pediatrics. 2005 Mar;115(3):662-6. doi: 10.1542/peds.2004-0820.
To evaluate usability, accuracy, and hypoglycemia detection of the GlucoWatch G2 Biographer (GW2B) in children aged 1 to 17 years.
After a 15-hour study of device accuracy, 46 children (15 <7 years, 31 > or =7 years) with type 1 diabetes were enrolled for an extended-wear outcome study: 2 daytime and 2 nighttime 15-hr wear periods each week and blood glucose monitoring 4 times daily for 3 months.
A total of 531 paired GW2B/meter readings were available for accuracy assessment. The correlation coefficients were 0.58 and 0.74 (ages <7 and > or =7 years, respectively). There was no significant change in hemoglobin A1C)or weight-adjusted insulin dose at 3 months after biographer use. Forty-two episodes of hypoglycemia were detected by the GW2B, 33 of which were confirmed by blood glucose meters. Sensitivity and specificity of audible low-glucose alerts were approximately 79% and 83%, respectively. No significant side effects were reported.
The GW2B is usable and safe in children who are <7 years or older in the home setting. The GW2B can detect asymptomatic nocturnal hypoglycemia in younger children.
评估GlucoWatch G2 Biographer(GW2B)在1至17岁儿童中的可用性、准确性及低血糖检测能力。
在对设备准确性进行15小时研究后,招募了46名1型糖尿病儿童(15名<7岁,31名≥7岁)进行长期佩戴结果研究:每周进行2个白天和2个夜间各15小时的佩戴期,并在3个月内每天进行4次血糖监测。
共有531对GW2B/血糖仪读数可用于准确性评估。相关系数分别为0.58和0.74(年龄<7岁和≥7岁组)。使用记录器3个月后,糖化血红蛋白或体重调整胰岛素剂量无显著变化。GW2B检测到42次低血糖发作,其中33次经血糖仪确认。可听式低血糖警报的敏感性和特异性分别约为79%和83%。未报告明显副作用。
GW2B在7岁以下或以上儿童的家庭环境中可用且安全。GW2B可检测年幼儿童无症状性夜间低血糖。