Higgs David, Uglow Michael, Fail Theresa
Queen Alexandra Hospital, Portsmouth.
J R Nav Med Serv. 2004;90(3):142-6.
The use of the A-V Impulse System of foot pumps and graduated compression stockings to reduce the incidence of thrombo-embolic disease after total hip arthroplasty is well recognised. It is not known if they have a synergistic effect as all clinical trials have used them in combination. We examined the effect compression stockings had on the ability of the A-V Impulse System to accelerate peak venous velocities in the common femoral vein in ten healthy volunteers using a duplex scanner. The use of foot pumps without stockings produced the greatest increase in peak venous velocity and this was 34% greater than using a foot pump with stockings. It is not known if this difference could account for a reduction in thrombo-embolic episodes and so we propose a randomised clinical trial.
使用动静脉脉冲系统的足部泵和分级压力弹力袜来降低全髋关节置换术后血栓栓塞性疾病的发生率已得到充分认可。由于所有临床试验都是将它们联合使用,因此尚不清楚它们是否具有协同作用。我们使用双功扫描仪检查了压力弹力袜对动静脉脉冲系统加快十名健康志愿者股总静脉峰值静脉速度能力的影响。不使用弹力袜时使用足部泵使峰值静脉速度增加最多,这比使用弹力袜时使用足部泵高出34%。尚不清楚这种差异是否能解释血栓栓塞事件的减少,因此我们提议进行一项随机临床试验。