[小腿经皮氧分压测量的可靠性]

[Reliability of transcutaneous measurement of oxygen tension on the lower leg].

作者信息

Käfer W, Kinkel S, Fraitzl C R, Clessienne C B, Puhl W, Huch K

机构信息

Orthopädische Klinik mit Querschnittgelähmtenzentrum der Universität Ulm.

出版信息

Z Orthop Ihre Grenzgeb. 2005 Jan-Feb;143(1):112-6. doi: 10.1055/s-2004-820346.

Abstract

AIM

Measurement of transcutaneous oxygen tension is increasingly used to determine the appropriate level of amputation in patients with vascular disease. The purpose of the present study was to analyze the intra- and interrater reliability of transcutaneous oxygen [tcpO (2)] measurements in a homogeneous study group.

METHOD

Five investigators assessed the transcutaneous oxygen tension of both lower legs of seven persons in a fixed setting. Assessment was repeated with the same examiners and the same examinees after 24 hours. TcpO (2) was measured at the posterior aspect of the lower leg twenty centimeters below the knee joint line. The TCM 400 Monitoring System (Radiometer Medical AIS, Bronshoj, Denmark) was used. Statistical analysis of the intra- and interrater reliability was performed with the Spearman coefficient of correlation.

RESULTS

An overall mean of 56.2 +/- 10.6 mmHg was found. For the first examination, a mean of 55.3 +/- 10.6 mmHg was observed, whereas for the second examination it was 57.0 +/- 10.5 mmHg. Analysis of intrarater reliability showed a coefficient of correlation of r (s) = 0.56 (p < 0.0001). For interrater reliability, we found coefficients of correlation ranging from r (s) = 0.20 (p = 0.20) to r (s) = 0.69 (p = 0.0004).

CONCLUSION

Analysis of transcutaneous oxygen tension measurements performed by different investigators in a fixed setting revealed a non- homogeneous intra- and interrater reliability, which should be taken into account prior to initiating therapy.

摘要

目的

经皮氧分压测量越来越多地用于确定血管疾病患者的合适截肢水平。本研究的目的是分析在一个同质研究组中经皮氧分压[tcpO₂]测量的观察者内和观察者间可靠性。

方法

五名研究者在固定环境下评估了七个人双侧小腿的经皮氧分压。24小时后,由相同的检查者和相同的受试者重复评估。在膝关节线以下20厘米处的小腿后侧测量tcpO₂。使用了TCM 400监测系统(丹麦布隆斯霍伊的Radiometer Medical AIS公司)。采用Spearman相关系数对观察者内和观察者间可靠性进行统计分析。

结果

总体平均值为56.2±10.6 mmHg。第一次检查时,观察到的平均值为55.3±10.6 mmHg,而第二次检查时为57.0±10.5 mmHg。观察者内可靠性分析显示相关系数r(s)=0.56(p<0.0001)。对于观察者间可靠性,我们发现相关系数范围从r(s)=0.20(p=0.20)到r(s)=0.69(p=0.0004)。

结论

对不同研究者在固定环境下进行的经皮氧分压测量分析显示,观察者内和观察者间可靠性不一致,在开始治疗前应予以考虑。