Suppr超能文献

目前牙科使用的18种不同X射线探测器的比较。

A comparison of 18 different x-ray detectors currently used in dentistry.

作者信息

Farman Allan G, Farman Taeko T

机构信息

Department of Surgical and Hospital Dentistry, The University of Louisville, Louisville, KY 40292, USA.

出版信息

Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2005 Apr;99(4):485-9. doi: 10.1016/j.tripleo.2004.04.002.

Abstract

PURPOSE

There has been a proliferation of available dental x-ray detectors over the recent past. The purpose of this short technical report is to provide a basic comparison of spatial resolution, contrast perceptibility, and relative exposure latitudes of 18 current dental x-ray detectors, including solid-state systems (CCD and CMOS), photostimulable phosphors, and analog film.

METHODS

Spatial resolution was measured using a 0.025 mm Pb phantom test grid with a measurement range from 1.5 to 20 lp/mm. For contrast perceptibility, a 7-mm thick aluminum perceptibility test device with wells of 0.1-0.9 mm depth at 0.1 mm intervals and 1 defect of 1.5 mm was used. The relative exposure latitude was determined by expert consensus using clear discrimination of the enamel-dentin junction as the lower limit and pixel blooming or unacceptable levels of cervical burnout as the upper limit.

RESULTS

The highest spatial resolution was found with Kodak InSight film, RVG-ui (CCD), and RVG 6000 (CMOS) detectors, each of which achieved 20 lp/mm, followed by the Planmeca Dixi2 v3 at > or =16 lp/mm. Contrast resolution was at least to 0.2 mm through 7 mm aluminum for all 18 detectors, with the best results being found for the Visualix HDI, RVG-ui, RVG 5000, and RVG 6000 detectors and the Schick CDR wired and wireless systems. The greatest exposure ranges were found with photostimulable phosphors and the Kodak RVG 6000 and RVG 5000 detectors.

CONCLUSIONS

Most current x-ray detectors generally perform well in terms of spatial and contrast resolutions, and in terms of exposure latitude. These findings were independent of the modality applied.

摘要

目的

近年来,可用的牙科X射线探测器数量激增。本简短技术报告的目的是对18种当前牙科X射线探测器的空间分辨率、对比度感知度和相对曝光宽容度进行基本比较,这些探测器包括固态系统(电荷耦合器件和互补金属氧化物半导体)、光激励荧光体和模拟胶片。

方法

使用0.025毫米铅制体模测试格栅测量空间分辨率,测量范围为1.5至20线对/毫米。对于对比度感知度,使用一个7毫米厚的铝制感知度测试装置,该装置有深度为0.1至0.9毫米、间隔为0.1毫米的孔以及一个1.5毫米的缺陷。相对曝光宽容度由专家共识确定,以清晰辨别釉牙本质界作为下限,以像素光晕或不可接受的颈部烧蚀水平作为上限。

结果

柯达InSight胶片、RVG-ui(电荷耦合器件)和RVG 6000(互补金属氧化物半导体)探测器的空间分辨率最高,均达到20线对/毫米,其次是普兰梅卡Dixi2 v3,其分辨率≥16线对/毫米。所有18种探测器在穿过7毫米铝时的对比度分辨率至少为0.2毫米,Visualix HDI、RVG-ui、RVG 5000和RVG 6000探测器以及施乐辉CDR有线和无线系统的结果最佳。光激励荧光体以及柯达RVG 6000和RVG 5000探测器的曝光范围最大。

结论

大多数当前的X射线探测器在空间和对比度分辨率以及曝光宽容度方面总体表现良好。这些发现与所应用的模式无关。