Takarabe Shinya, Kuramoto Taku, Shibayama Yusuke, Tsuru Hiroki, Tatsumi Masato, Kato Toyoyuki, Okamura Kazutoshi, Yoshiura Kazunori
Kyushu University Hospital, Division of Radiology, Department of Medical Technology, Fukuoka, Japan.
Kyushu University, Department of Oral and Maxillofacial Radiology, Faculty of Dental Science, Fukuoka, Japan.
J Med Imaging (Bellingham). 2021 Jul;8(4):043501. doi: 10.1117/1.JMI.8.4.043501. Epub 2021 Jul 20.
The objective of this study was to investigate the effects of (i) the difference in the beam qualities on the presampled modulation transfer function (MTF) using the edge method and (ii) the readout direction of the edge profile for the photostimulable phosphor (PSP) system. The International Electrotechnical Commission (IEC) defined a technique using the "radiation qualities based on a phantom made up of an aluminum added filter" (RQA). A general radiographic system with a tube voltage of 50 kV and 9.7 mm of additional aluminum filtration was used to conform the x-ray to the IEC-specified beam quality definition RQA3. Additionally, we employed two different beams with tube voltages of 60 and 70 kV using a dental x-ray unit. The MTF was measured in the readout direction from low-to-high exposure regions and vice versa with respect to the scanning and subscanning directions of the PSP system. The difference in the 50%MTF value for all directions between 60 and 70 kV averaged less than . The 50%MTF of RQA3 was on average 0.2 lower than the value for 60 kV for each direction. For all beam qualities, no difference was observed between the MTFs measured in the readout direction from the low-to-high exposure regions and vice versa. The MTFs, measured using the dental x-ray unit, were unaffected by the tube voltage, and they were slightly higher than those measured using the RQA3. Furthermore, the MTF was unaffected by the differences in the readout directions of the edge profile.
(i) 使用边缘法时,不同射线质对预采样调制传递函数 (MTF) 的影响;(ii) 光激励荧光体 (PSP) 系统边缘轮廓读出方向的影响。国际电工委员会 (IEC) 定义了一种使用 “基于添加铝滤过体的体模的辐射质” (RQA) 的技术。使用管电压为50 kV且附加铝滤过为9.7 mm的普通射线照相系统,使X射线符合IEC规定的射线质定义RQA3。此外,我们使用牙科X射线机,采用管电压分别为60 kV和70 kV的两种不同射线。MTF是在PSP系统的扫描和副扫描方向上,从低曝光区域到高曝光区域以及相反方向的读出方向上进行测量的。60 kV和70 kV之间所有方向的50%MTF值差异平均小于 。RQA3的50%MTF在每个方向上平均比60 kV的值低0.2。对于所有射线质,在从低曝光区域到高曝光区域以及相反方向的读出方向上测量的MTF之间未观察到差异。使用牙科X射线机测量的MTF不受管电压影响,且略高于使用RQA3测量的MTF。此外,MTF不受边缘轮廓读出方向差异的影响。