Della Barbera Mila, Laborde François, Thiene Gaetano, Arata Valter, Pettenazzo Elena, Pasquino Enrico, Behr Luc, Valente Marialuisa
Institute of Pathological Anatomy, University of Padua Medical School, Padua, Italy.
Cardiovasc Pathol. 2005 Mar-Apr;14(2):96-103. doi: 10.1016/j.carpath.2004.12.005.
A new flexible annuloplasty ring (Sovering, Sorin Biomedica Cardio), both closed and open, has been designed and consists of radiopaque silicone core impregnated with barium sulfate and covered by a knitted polyester fabric coated with Carbofilm. The aim of the study was to test the biological compatibility of the new device in large animals in orthotopic position.
Ten Sovering rings were implanted in 10 female adult sheep, 7 mitral (3 open, 4 closed) and 3 tricuspid (all open). The size was 23-25 mm in the mitral and 28-30 mm in the tricuspid position, and the time in place varied from 63 to 110 days (mean = 89 +/- 14) and from 58 to 63 days (mean = 61 +/- 3), respectively. The morphological analysis consisted of gross, histological, immunohistochemical and ultrastructural investigations.
The prosthetic ring appeared well implanted in the valve atrioventricular (AV) junction, encapsulated by a thin fibrous sheath without any evidence of thrombus deposition, fibrinous lining or exuberant fibrous tissue overgrowth. No adverse inflammatory reaction was observed, but rare lymphocytes, macrophages and foreign body giant cells. At electron microscopy, the fibrous tissue appeared to permeate deeply within the fabric network; reendothelization on the surface was noted and confirmed by immunohistochemistry. Sirius red staining at polarized light revealed a higher content of collagen type III in the mitral than in the tricuspid position.
Sovering annuloplasty rings implanted in the AV valves of adult sheep showed excellent biocompatibility, fibrous encapsulation and reendothelization. The absence of thrombosis and exuberant fibrous tissue reaction supports the effective use of Carbofilm covering in annuloplasty devices.
一种新型的可弯曲瓣环成形环(Sovering,索林生物医学心脏公司),有闭合和开放两种类型,其设计为由硫酸钡浸渍的不透射线硅胶芯和涂有碳膜的针织聚酯织物覆盖层组成。本研究的目的是在大动物的原位测试该新装置的生物相容性。
将10个Sovering环植入10只成年雌性绵羊体内,7个用于二尖瓣(3个开放型,4个闭合型),3个用于三尖瓣(均为开放型)。二尖瓣位置的尺寸为23 - 25毫米,三尖瓣位置为28 - 30毫米,植入时间分别为63至110天(平均 = 89 ± 14天)和58至63天(平均 = 61 ± 3天)。形态学分析包括大体、组织学、免疫组织化学和超微结构研究。
人工瓣环在房室(AV)交界处的瓣膜中植入良好,被一层薄纤维鞘包裹,没有血栓沉积、纤维蛋白内衬或过度增生的纤维组织的迹象。未观察到不良炎症反应,但有罕见的淋巴细胞、巨噬细胞和异物巨细胞。在电子显微镜下,纤维组织似乎深入渗透到织物网络中;表面有内皮化现象,并经免疫组织化学证实。偏振光下的天狼星红染色显示二尖瓣位置的III型胶原蛋白含量高于三尖瓣位置。
植入成年绵羊AV瓣膜的Sovering瓣环成形环显示出优异的生物相容性、纤维包裹和内皮化。无血栓形成和过度增生的纤维组织反应支持了在瓣环成形装置中有效使用碳膜覆盖层。