Pulpal and periodontal temperature rise during KTP laser use as a root planing complement in vitro.

作者信息

Nammour S, Rocca J-P, Keiani K, Balestra C, Snoeck T, Powell L, Reck J Van

机构信息

Department of Dental Sciences, Faculty of Medicine, University of Liege, Liege, Belgium.

出版信息

Photomed Laser Surg. 2005 Feb;23(1):10-4. doi: 10.1089/pho.2005.23.10.

Abstract

OBJECTIVE

The purpose of this study was to define the optimal irradiation conditions of a KTP laser during root planing treatment.

METHODS

The surfaces of 60 single-root human teeth were scaled with conventional instruments before lasing. The pulpal temperature increase was measured by means of one thermocouple placed in the pulp chamber and a second one placed on the root surface at 1 mm from the irradiation site. The influence of variables of coloration by Acid Red 52 (photosensitizer), scanning speed, dentin thickness, and probe position was analyzed for a constant exposure time of 15 sec and 500 mw (spot size diameter, 0.5 mm). The pulpal temperature was below 3 degrees C for the adjustments.

RESULTS

The irradiation on one point of root surface had the following results: The application of photosensitizer on the root surface before lasing produced a 50% higher temperature rise within the pulp than in the case without the application of the photosensitizer. The temperature rise in the pulp chamber was below 3 degrees C with the following settings of 500 mw: PW = 10 msec and PRR < 35; or PW= 20 msec and PRR < 20 Hz. On the other hand, for the same irradiation conditions, the temperature rise on the surface of the root was always below 7 degrees C. However, the temperature increase became higher than 7 degrees C (on the surface of the root) in the case of P > 500 mw, PW > 50 msec and PRR > 10 Hz of root surface or a scanning speed of irradiation of 1 mm/sec for a linear irradiation of 4 mm.

CONCLUSION

The KTP laser may be used safely without thermal damage to pulp and periodontal tissue with respect to the biologically acceptable previously described parameters.

摘要

目的

本研究的目的是确定在根面平整治疗过程中KTP激光的最佳照射条件。

方法

60颗单根人牙的表面在激光照射前用传统器械进行刮治。通过放置在髓腔内的一个热电偶和放置在距照射部位1毫米处根面上的另一个热电偶来测量牙髓温度的升高。在15秒恒定曝光时间和500毫瓦(光斑尺寸直径,0.5毫米)的条件下,分析了酸性红52(光敏剂)染色、扫描速度、牙本质厚度和探头位置等变量的影响。调整后的牙髓温度低于3摄氏度。

结果

在根面一点进行照射有以下结果:激光照射前在根面上应用光敏剂,牙髓内温度升高比未应用光敏剂的情况高50%。在500毫瓦的以下设置下,髓腔内温度升高低于3摄氏度:脉宽(PW)=10毫秒且脉冲重复率(PRR)<35;或PW = 20毫秒且PRR < 20赫兹。另一方面,对于相同的照射条件,根面温度升高始终低于7摄氏度。然而,在根面功率(P)> 500毫瓦、脉宽> 50毫秒且脉冲重复率> 10赫兹或线性照射4毫米时照射扫描速度为1毫米/秒的情况下,根面温度升高超过7摄氏度。

结论

就先前所述的生物学可接受参数而言,KTP激光可以安全使用,不会对牙髓和牙周组织造成热损伤。