Yaneva Blagovesta K, Zagorchev Plamen I, Firkova Elena I, Glavinkov Ivan T
Department of Periodontics and Oral Mucosa Diseases, Faculty of Dental Medicine, Medical University of Plovdiv, Plovdiv, Bulgaria
Department of Medical Physics, Biophysics and Mathematics, Faculty of Pharmacy, Medical University of Plovdiv, Plovdiv, Bulgaria
Folia Med (Plovdiv). 2016 Sep 1;58(3):206-210. doi: 10.1515/folmed-2016-0022.
To assess temperature changes at specified time intervals during Er:YAG laser scaling and root planing of surfaces with dental calculus.
Fifteen single-rooted teeth with advanced periodontal disease were extracted and fixed in a cylinder thermostat filled with distilled water at constant temperature (35.5°C). A specially designed thermal probe (type K thermocouple) accurate to ±0.1°C over the range from 20°C to 80°C was fitted into the pulp chamber of tooth sample. Scaling and root planing of the mesial and distal root surfaces was performed using an Er:YAG laser (Lite Touch, Syneron Dental, Israel) with a wavelength of 2940 nm, provided with a chisel tip, and at the following settings: output energy 100 mJ and 50 Hz, duration of irradiation - 40 sec, the tip in contact mode oblique to the root surface at an angle of approximately 10-15 degrees and water spray level 5-6. The temperature inside the pulp chamber was measured every 10 sec.
The temperature in the pulp chamber taken every 10 seconds and compared with the temperature of 35.5°C at baseline decreased by 1.6°C, 2.4°C, 2.5°C, and 2.5°C for the first, second, third and fourth measurement, respectively. These changes did not reach statistical significance.
The Er:YAG laser does not increase the temperature inside the pulp chamber. The assessed changes do not depend on the duration of irradiation which was kept within 40 seconds. Therefore, this treatment modality causes no thermal damage to the pulp under the above defined conditions and can be considered safe.
评估在使用铒激光对有牙结石的牙面进行洁治和根面平整过程中特定时间间隔的温度变化。
选取15颗患有晚期牙周病的单根牙,将其拔除并固定于一个装有恒温(35.5°C)蒸馏水的圆柱恒温器中。将一个在20°C至80°C范围内精度为±0.1°C的特制热探头(K型热电偶)安装到牙齿样本的髓腔中。使用波长为2940nm、配有凿形尖端的铒激光(Lite Touch,Syneron Dental,以色列)对近中根面和远中根面进行洁治和根面平整,设置如下:输出能量100mJ、频率50Hz、照射持续时间40秒、尖端以约10 - 15度角倾斜于根面的接触模式以及水喷雾水平5 - 6。每10秒测量一次髓腔内的温度。
每10秒测量一次的髓腔内温度与基线时35.5°C的温度相比,第一次、第二次、第三次和第四次测量分别下降了1.6°C、2.4°C
、2.5°C和2.5°C。这些变化未达到统计学显著性。
铒激光不会增加髓腔内的温度。所评估的变化不取决于保持在40秒内的照射持续时间。因此,在上述限定条件下,这种治疗方式不会对牙髓造成热损伤,可被认为是安全的。