Garcia Julian P S, Garcia Patricia M T, Rosen Richard B, Finger Paul T
The New York Eye & Ear Infirmary, New York, New York 10003, USA.
Ophthalmic Surg Lasers Imaging. 2005 Mar-Apr;36(2):142-6.
Twenty-three normal eyes were examined with the Intensity Profiling technique of 3D I-Scan and 52 normal eyes with the Automated technique of OTI-Scan 1000 ultrasound systems (Ophthalmic Technologies Inc., Toronto, Ontario, Canada). With the eye looking straight, the probe was applied on the temporal sclera. Scanning generated 3D image files. Coronal optic nerve measurements were obtained 3 mm behind the globe. The mean optic nerve sheath diameter was 4.8 mm (standard deviation = 0.6; range, 3.9 to 5.9 mm) with the Intensity Profiling technique, and 5.4 mm (standard deviation = 0.4; range, 4.4 to 6.0 mm) with the Automated technique. 3D ultrasound imaging can be used to obtain optic nerve measurements in vivo. Both the Intensity Profiling and the Automated techniques yielded measurements similar to current magnetic resonance imaging and computed tomography scan reports. Coronal "C-scan" sectioning can be used as a screening tool to measure optic nerve diameter prior to magnetic resonance imaging or computed tomography.
使用3D I-Scan的强度分析技术对23只正常眼睛进行了检查,使用OTI-Scan 1000超声系统(加拿大安大略省多伦多市眼科技术公司)的自动技术对52只正常眼睛进行了检查。眼睛直视时,将探头置于颞侧巩膜上。扫描生成3D图像文件。在眼球后方3毫米处获得冠状位视神经测量值。强度分析技术测得的平均视神经鞘直径为4.8毫米(标准差=0.6;范围为3.9至5.9毫米),自动技术测得的平均视神经鞘直径为5.4毫米(标准差=0.4;范围为4.4至6.0毫米)。3D超声成像可用于在体内获取视神经测量值。强度分析技术和自动技术得出的测量结果与当前磁共振成像和计算机断层扫描报告相似。冠状位“C扫描”切片可作为在磁共振成像或计算机断层扫描之前测量视神经直径的筛查工具。