Suppr超能文献

用于功能性电刺激移动系统的定向传感器的性能

Performance of orientation sensors for use with a functional electrical stimulation mobility system.

作者信息

Simcox Scott, Parker Simon, Davis Glen M, Smith Richard W, Middleton James W

机构信息

Rehabilitation Research Centre, The University of Sydney, Australia.

出版信息

J Biomech. 2005 May;38(5):1185-90. doi: 10.1016/j.jbiomech.2004.05.026.

Abstract

The purpose of this study was to verify the performance of recently developed body-worn sensor packs against 3D motion analysis of trunk and lower-limb movements. Five sensor packs, each consisting of rate gyroscope and two 2-D accelerometers controlled by a microprocessor were attached to the trunk, thighs, and shanks of an able bodied subject. A 6-camera motion analysis system (MAS) recorded multiple trials of sit-to-stand movements and normal walking. Time domain signals from each sensor pack were significantly correlated (r = 0.90-0.99;p < 0.05) with a root mean square errors of less than 5 degrees when compared against the same limb angle measurements calculated by the MAS. These data demonstrate that these external sensor packs are accurate devices for measuring trunk and lower-limb sagittal plane orientation in real-time.

摘要

本研究的目的是验证最近开发的可穿戴式传感器组在测量躯干和下肢运动的三维运动分析方面的性能。五个传感器组,每个由速率陀螺仪和两个由微处理器控制的二维加速度计组成,被附着在一名身体健全受试者的躯干、大腿和小腿上。一个六摄像头运动分析系统(MAS)记录了多次从坐姿到站姿运动和正常行走的试验。与MAS计算出的相同肢体角度测量值相比,每个传感器组的时域信号具有显著相关性(r = 0.90 - 0.99;p < 0.05),均方根误差小于5度。这些数据表明,这些外部传感器组是实时测量躯干和下肢矢状面方向的精确设备。