使用微传感器进行不同颅内压(ICP)监测途径的准确性。

Accuracy of different routes of intracranial pressure (ICP) monitoring using microtransducers.

作者信息

Poon W S, Li A K

机构信息

Neurosurgical Unit, Department of Surgery, Prince of Wales Hospital, Chinese University of Hong Kong, Shatin, Hong Kong.

出版信息

J Neurosurg Anesthesiol. 1990 Dec;2(4):280-3. doi: 10.1097/00008506-199012000-00005.

Abstract

Intracranial pressure (ICP) monitoring using microtransducer and intraventricular pressure (IVP) was compared simultaneously in 20 consecutive head-injured patients within a 1-year period. In the first four patients with subdurally placed transducers, an acceptable corresponding rate with IVP was demonstrated (88%). In the remaining 16 patients with extradurally placed transducers, acceptable accuracy (73%) was only achieved in the last four. This accuracy was repeated in an additional six patients with extradurally placed transducers (75% corresponding with IVP). We have therefore suggested that, before extradural pressure monitoring can be employed alone in clinical practice, a comparison with IVP should be performed.

摘要

在1年时间内,对20例连续的头部受伤患者同时比较了使用微型传感器进行的颅内压(ICP)监测和脑室内压力(IVP)。在前4例硬膜下放置传感器的患者中,显示出与IVP可接受的对应率(88%)。在其余16例硬膜外放置传感器的患者中,仅在最后4例中达到了可接受的准确性(73%)。另外6例硬膜外放置传感器的患者也重复了这一准确性(与IVP的对应率为75%)。因此,我们建议,在临床实践中单独采用硬膜外压力监测之前,应与IVP进行比较。