Evaluating cochlear implant trauma to the scala vestibuli.

作者信息

Adunka O, Kiefer J, Unkelbach M H, Radeloff A, Gstoettner W

机构信息

Clinic of Otolaryngology, Head and Neck Surgery, Johann Wolfgang Goethe University, Frankfurt am Main, Germany.

出版信息

Clin Otolaryngol. 2005 Apr;30(2):121-7. doi: 10.1111/j.1365-2273.2004.00935.x.

Abstract

OBJECTIVES

Placement of cochlear implant electrodes into the scala vestibuli may be intentional, e.g. in case of blocked scala tympani or unintentional as a result of trauma to the basilar membrane or erroneous location of the cochieostomy. The aim of this study was to evaluate the morphological consequences and cochlear trauma after implantation of different cochlear implant electrode arrays in the scala vestibuli.

DESIGN

Human temporal bone study with histological and radiological evaluation.

SETTING

Twelve human cadaver temporal bones were implanted with different cochlear implant electrodes. Implanted bones were processed using a special method to section undecalcified bone.

MAIN OUTCOME MEASURES

Cochlear trauma and intracochlear positions.

RESULTS

All implanted electrodes were implanted into the scala vestibuli using a special approach that allows direct scala vestibuli insertions. Fractures of the osseous spiral lamina were evaluated in some bones in the basal cochlear regions. In most electrodes, delicate structures of the organ of Corti were left intact, however, Reissner's membrane was destroyed in all specimens and the electrode lay upon the tectorial membrane. In some bones the organ of Corti was destroyed.

CONCLUSIONS

Scala vestibuli insertions did not cause severe trauma to osseous or neural structures, thus preserving the basis for electrostimulation of the cochlea. However, destruction of Reissner's membrane and impact on the Organ of Corti can be assumed to destroy residual hearing.

摘要

目的

将人工耳蜗电极植入前庭阶可能是有意为之,例如在鼓阶堵塞的情况下,或者是由于基底膜创伤或耳蜗造口位置错误而意外发生。本研究的目的是评估在前庭阶植入不同人工耳蜗电极阵列后的形态学后果和耳蜗损伤情况。

设计

采用组织学和放射学评估的人体颞骨研究。

设置

对12个尸体颞骨植入不同的人工耳蜗电极。使用特殊方法处理植入的骨头以切片未脱钙骨。

主要观察指标

耳蜗损伤和耳蜗内位置。

结果

所有植入电极均采用一种允许直接插入前庭阶的特殊方法植入前庭阶。在一些基底耳蜗区域的骨头中评估了骨螺旋板的骨折情况。在大多数电极中,柯蒂氏器的精细结构保持完整,然而,所有标本中的Reissner膜均被破坏,电极位于盖膜上。在一些骨头中,柯蒂氏器被破坏。

结论

前庭阶插入不会对骨结构或神经结构造成严重损伤,从而保留了耳蜗电刺激的基础。然而,可以认为Reissner膜的破坏和对柯蒂氏器的影响会破坏残余听力。