Implantation of permanent pacemaker systems is associated with a low rate of complications. The case of a 72-year-old woman is described who suffered from a cluster of complications related to permanent pacemaker implantation. Besides major bleeding, requiring transfusion and pneumothorax, malposition of both pacemaker leads in the arterial system occurred, with the atrial lead screwed into the right coronary artery, with subsequent occlusion of the vessel and the ventricle lead placed in the left ventricle.
永久性心脏起搏器系统植入术的并发症发生率较低。本文描述了一名72岁女性患者,她出现了一系列与永久性心脏起搏器植入相关的并发症。除了需要输血的大出血和气胸外,两根起搏器导线在动脉系统中均发生位置异常,心房导线拧入右冠状动脉,随后该血管闭塞,心室导线置于左心室。