Nie Bang-ji, Xu Long, Xin Xue-gang, Wang Cheng-lai, Wu Min-shan
Dept. of Biomedical Engineering, Sun Yat-Sen University, Guangzhou.
Zhongguo Yi Liao Qi Xie Za Zhi. 2005 Jan;29(1):5-6.
Some research on intelligent non-invasive temporary pacemakers is introduced in this paper. An industrial computer, some IC chips and other elements are used to construct its hardware, and its software is in C++ language. The experimental device has some intelligent functions of recognizing some arrhythmia. The system has a pacemaker module and an ECG monitor module. Its software includes a main program, a RS-232C communication program, a printer VxD, a pacing control VxD and ECG signal pretreatment and recognizing program and so on. The pacing-generating circuit is employed to make the precision control of pacing current. The communication between industrial-computer system and ECG module is completed through the DLL. The real time processing of ECG signals is based on filter method for a higher recognizing ratio. The system calculates several parameters to recognize certain arrhythmia and uses MIT/BIH database to validate the reliability of ECG recognition.