Sakama Makoto, Kanai Tatsuaki, Kase Yuki, Komori Masataka, Fukumura Akifumi, Kohno Toshiyuki
Tokyo Institute of Technology, 4259, Nagatsuta-cho, Midori-ku, Yokohama 226-8502, Japan.
Phys Med Biol. 2005 May 21;50(10):2275-89. doi: 10.1088/0031-9155/50/10/007. Epub 2005 Apr 27.
The responses of a commercial diamond detector (type 60003, PTW-Freiburg) to several heavy ions were examined. The responses to heavy-ion beams reached stable levels with relatively small pre-irradiation doses compared to photon-beam irradiations. The responses also reached stable levels with a smaller pre-irradiation dose when the dose rate of the He beams was increased. A total accumulated dose of about 5 Gy was required for the pre-irradiation dose of heavy-ion beams. No angular dependence of the detector responses was observed within a deviation of 5%. The dose-rate dependence of the detector responses to heavy-ion beams was far smaller than that to gamma rays. The decrease in the response was within 0.9%, with a variation from 0.88 to 18.2 Gy min(-1) in the carbon beam. We examined the LET dependence of the diamond detector responses using various kinds of heavy-ion beams. The responses had particle dependence in addition to LET dependence. The responses decreased more with higher LET particles and decreased less with large-Z particles. We proposed a gradual-saturation model based on the track structure under several simple assumptions to explain the LET and particle dependences of the response.
研究了一种商用金刚石探测器(型号60003,PTW - 弗莱堡)对几种重离子的响应。与光子束辐照相比,重离子束辐照在相对较小的预辐照剂量下响应就达到了稳定水平。当氦束的剂量率增加时,预辐照剂量较小时响应也能达到稳定水平。重离子束的预辐照剂量需要约5 Gy的总累积剂量。在5%的偏差范围内未观察到探测器响应的角度依赖性。探测器对重离子束响应的剂量率依赖性远小于对伽马射线的依赖性。在碳束中,剂量率从0.88到18.2 Gy min⁻¹变化时,响应的下降幅度在0.9%以内。我们使用各种重离子束研究了金刚石探测器响应的传能线密度(LET)依赖性。响应除了具有LET依赖性外还具有粒子依赖性。LET较高的粒子使响应下降得更多,而原子序数(Z)较大的粒子使响应下降得较少。我们基于几个简单假设下的径迹结构提出了一个渐变饱和模型来解释响应的LET和粒子依赖性。