Levitt M A, Lopez B, Lieberman M E, Sutton M
Division of Emergency Medicine, Thomas Jefferson University Hospital, Philadelphia, Pennsylvania.
Ann Emerg Med. 1992 Jun;21(6):713-8. doi: 10.1016/s0196-0644(05)82785-4.
To evaluate the usefulness of the tilt test in identifying significantly dehydrated or occultly bleeding adult patients in an emergency department.
Prospective.
Urban ED.
Two hundred two ill adult patients with complaints suggestive of dehydration and/or occult blood loss and a control sample of 21 healthy individuals.
Orthostatic measurements were taken from blood samples from the ill and healthy patients. The estimates of the degree of dehydration measured or calculated were total body water deficit percentage (based on measured serum osmolality), ratio of blood urea nitrogen to creatinine, creatinine, and hematocrit. Multiple historical and physical examination variables related to dehydration or occult bleeding were recorded.
Multiple analysis of variance testing of the historical and examination variables in the 166 dehydrated patients revealed the absence of syncope (P = .037) and the lack of axillary sweat (P = .026) to be significantly associated with an increased level of dehydration. Regression analysis of the orthostatic measurements and age with the estimates of dehydration found that orthostatic change in heart rate (P = .0165) and age (P = .0047) demonstrated a very small (r2 = .098) association with the level of dehydration. Orthostatic changes in systolic blood pressure and diastolic blood pressure demonstrated no statistically significant association with level of dehydration. There was wide variation in the tilt test measurements in samples from both the healthy and the ill patients: heart rate, 6.8 +/- 7.8 versus 13.2 +/- 10.5; systolic blood pressure, -2.5 +/- 8.0 versus -8.3 +/- 12.8 mm Hg; and diastolic blood pressure, 5.3 +/- 9.9 versus 0.6 +/- 8.6 mm Hg, respectively. The only parameter that attained statistical significance between the 36 blood-loss patients and the healthy patients was change in systolic blood pressure (P = .001); however, this was not clinically significant (-10.7 +/- 13.7 versus -2.5 +/- 8.0 mm Hg, respectively).
It appears impossible to define a "positive" tilt test that would adequately identify patients with clinically significant dehydration or blood loss; this is due to the large variance in patients' orthostatic measurements both in a healthy and in an ill state and the lack of a significant correlation of orthostatic measurements to a level of dehydration. Lack of axillary sweat, no complaint of syncope, and younger age all indicate greater degrees of dehydration.
评估倾斜试验在急诊科鉴别严重脱水或隐匿性出血成年患者中的作用。
前瞻性研究。
城市急诊科。
202例有脱水和/或隐匿性失血症状的成年患者以及21名健康个体作为对照样本。
对患病和健康患者的血样进行直立位测量。测量或计算得到的脱水程度估计指标包括总体液缺失百分比(基于测得的血清渗透压)、血尿素氮与肌酐比值、肌酐和血细胞比容。记录与脱水或隐匿性出血相关的多个病史和体格检查变量。
对166例脱水患者的病史和检查变量进行多因素方差分析,结果显示无晕厥(P = 0.037)和无腋窝出汗(P = 0.026)与脱水程度增加显著相关。对直立位测量指标、年龄与脱水程度估计值进行回归分析发现,心率的直立位变化(P = 0.0165)和年龄(P = 0.0047)与脱水程度呈非常小的相关性(r2 = 0.098)。收缩压和舒张压的直立位变化与脱水程度无统计学显著相关性。健康和患病患者样本的倾斜试验测量值差异很大:心率分别为6.8±7.8与13.2±10.5;收缩压分别为-2.5±8.0与-8.3±12.8 mmHg;舒张压分别为5.3±9.9与0.6±8.6 mmHg。36例失血患者与健康患者之间唯一达到统计学显著意义的参数是收缩压变化(P = 0.001);然而,这在临床上并不显著(分别为-10.7±13.7与-2.5±8.0 mmHg)。
似乎无法定义一个能充分鉴别具有临床显著脱水或失血患者的“阳性”倾斜试验;这是由于健康和患病状态下患者直立位测量值差异很大,且直立位测量值与脱水程度缺乏显著相关性。无腋窝出汗、无晕厥主诉和年龄较小均提示脱水程度较重。