Suppr超能文献

毛细管电泳-质谱联用中的定量分析:长期和短期方差分量及其使用内标的补偿

Quantification in capillary electrophoresis-mass spectrometry: long- and short-term variance components and their compensation using internal standards.

作者信息

Ohnesorge Jens, Sänger-van de Griend Cari, Wätzig Hermann

机构信息

Institut für Pharmazeutische Chemie, TU Braunschweig, Braunschweig, Germany.

出版信息

Electrophoresis. 2005 Jun;26(12):2360-75. doi: 10.1002/elps.200410413.

Abstract

Different approaches were chosen to examine ionization reproducibility of analytes after separation by capillary electrophoresis-mass spectrometry (CE-MS) in a commercially available sheath-flow electrospray interface. For this task three different standard samples were examined. Sample 1 contained neostigmine bromide (cationic), paracetamol (PCM) (neutral) and nicotinic acid (anionic component). Results were evaluated using internal standard (IS) calculations. Sample 2 represented an isotopically labelled IS of the quantified substance (PCM/D4-PCM), while sample 3 (neostigmine bromide/scopolamine hydrobromide) provided an IS closely migrating to the tested substance. Furthermore, short-time variations inside the interface were examined by multiple injections of the same substance. For sample 1, the relative standard deviations (RSD%s) were between 8 and 25% (n at least 58) for the peak area ratios. Multiple injected samples gave 5.5-19.4% (n = 25) for peak area RSD%. Using a closely migrating IS, sample 3, RSD%s between 6.5 and 10% (n at least 63) were achieved. With isotopically labelled IS, sample 2, an RSD% of 3-4% was achieved for peak area ratios over long periods (n = 25), for shorter periods (n = 9) even 1-2% RSD% was obtained. Keeping the instrument settings constant, the influence on the ionization efficiency and reproducibility was tested, varying the buffer pH, the organic buffer modifier and the sample concentration. Repeatabilities of migration time and peak area were measured and compared. Two 10 mM ammonium acetate buffers with pH 4.0 and 8.5 were investigated. No influence of buffer pH on peak area reproducibility was found. Isopropanol as organic buffer modifier significantly improved the ionisation leading to larger peak areas, but reduced reproducibility. The basic buffer produced slightly better RSD%s for migration times (2.5-4.0%) (n = 180) and faster analysis for the different test analytes of sample 1, while with the acetic buffer, RSD%s from 3.9 to 6.0% were obtained (n at least 163). The positioning of the capillary turned out to be the crucial parameter to ensure reproducible results. Thus, a procedure was established to ensure a defined ion-intensity level after capillary changes. The investigation of the different sample concentrations gave negligible differences in RSD%, showing that the signal-to-noise ratio was not the crucial parameter for reproducibility here, in contrast to CE-UV detection.

摘要

采用不同方法研究了在市售鞘流电喷雾接口中通过毛细管电泳-质谱联用(CE-MS)分离后分析物的电离重现性。为此任务检测了三种不同的标准样品。样品1包含溴新斯的明(阳离子)、对乙酰氨基酚(PCM)(中性)和烟酸(阴离子成分)。结果采用内标(IS)计算进行评估。样品2代表定量物质的同位素标记内标(PCM/D4-PCM),而样品3(溴新斯的明/氢溴酸东莨菪碱)提供了一种迁移行为与被测物质相近的内标。此外,通过多次进样同一种物质研究了接口内部的短期变化。对于样品1,峰面积比的相对标准偏差(RSD%)在8%至25%之间(n至少为58)。多次进样的样品峰面积RSD%为5.5%至19.4%(n = 25)。使用迁移行为相近的内标(样品3),RSD%在6.5%至10%之间(n至少为63)。对于同位素标记的内标(样品2),长时间(n = 25)峰面积比的RSD%达到3%至4%,短时间(n = 9)时甚至获得1%至2%的RSD%。在保持仪器设置不变的情况下,通过改变缓冲液pH值、有机缓冲液改性剂和样品浓度,测试了对电离效率和重现性的影响。测量并比较了迁移时间和峰面积的重复性。研究了两种pH值分别为4.0和8.5的10 mM醋酸铵缓冲液。未发现缓冲液pH值对峰面积重现性有影响。异丙醇作为有机缓冲液改性剂显著提高了电离,导致峰面积增大,但降低了重现性。碱性缓冲液在迁移时间方面产生的RSD%略好(2.5%至4.0%)(n = 180),且对样品1的不同测试分析物分析速度更快,而使用醋酸缓冲液时,RSD%为3.9%至6.0%(n至少为163)。结果表明,毛细管的定位是确保结果可重现的关键参数。因此,建立了一种程序,以确保毛细管更换后有确定的离子强度水平。对不同样品浓度的研究表明,RSD%的差异可忽略不计,这表明与CE-UV检测不同,此处信噪比不是重现性的关键参数。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验