Medici G, Mussi C, Fantuzzi A L, Malavolti M, Albertazzi A, Bedogni G
Cattedra di Nefrologia, Università di Modena e Reggio Emilia, Italy.
Eur J Clin Nutr. 2005 Aug;59(8):932-7. doi: 10.1038/sj.ejcn.1602165.
To establish the accuracy of bioelectrical impedance analysis (BIA) for the assessment of total and appendicular body composition in peritoneal dialysis (PD) patients.
Cross-sectional study.
University Nephrology Clinic.
In all, 20 PD patients and 77 healthy controls matched for gender, age and body mass index.
Whole-body fat-free mass (FFM) and appendicular lean tissue mass (LTM) were measured by dual-energy X-ray absorptiometry. Resistance (R) of arms, trunk and legs was measured by eight-polar BIA at frequencies of 5, 50, 250 and 500 kHz. Whole-body resistance was calculated as the sum of R of arms, trunk and legs. The resistance index (RI) was calculated as the ratio between squared height and whole-body or segmental R.
RI at 500 kHz was the best predictor of FFM, LTM(arm) and LTM(leg) in both PD patients and controls. Equations developed on controls overestimated FFM and LTM(arm) and underestimated LTM(leg) when applied to PD patients. Specific equations were thus developed for PD patients. Using these equations, the percent root mean-squared errors of the estimate for PD patients vs controls were 5 vs 6% for FFM, 8 vs 8% for LTM(arm) and 7 vs 8% for LTM(leg).
Eight-polar BIA offers accurate estimates of total and appendicular body composition in PD patients, provided that population-specific equations are used.
确定生物电阻抗分析(BIA)评估腹膜透析(PD)患者总体和附属身体成分的准确性。
横断面研究。
大学肾脏病诊所。
共20例PD患者和77名性别、年龄和体重指数相匹配的健康对照者。
采用双能X线吸收法测量全身去脂体重(FFM)和附属瘦组织质量(LTM)。通过八极生物电阻抗分析在5、50、250和500kHz频率下测量手臂、躯干和腿部的电阻(R)。全身电阻计算为手臂、躯干和腿部电阻之和。电阻指数(RI)计算为身高平方与全身或节段电阻的比值。
在PD患者和对照组中,500kHz时的RI是FFM、LTM(手臂)和LTM(腿部)的最佳预测指标。基于对照组建立的方程应用于PD患者时,高估了FFM和LTM(手臂),低估了LTM(腿部)。因此为PD患者建立了特定方程。使用这些方程,PD患者与对照组估计值的均方根误差百分比,FFM分别为5%和6%,LTM(手臂)分别为8%和8%,LTM(腿部)分别为7%和8%。
如果使用针对特定人群的方程,八极生物电阻抗分析能够准确估计PD患者的总体和附属身体成分。