Problems with film processing in medical X-ray imaging in Lithuania.

作者信息

Sniureviciute Migle, Adliene Diana

机构信息

Radiation Protection Centre, Kaunas department, Kaunas, Lithuania.

出版信息

Radiat Prot Dosimetry. 2005;114(1-3):260-3. doi: 10.1093/rpd/nch577.

Abstract

Optimisation in X-ray imaging in order to reduce patient doses during diagnostic X-ray examinations is a complex process given the high level of image quality required. When quality systems are implemented as a basis for optimisation, attention should be paid to the qualifications of the staff and quality control of the equipment and of the X-ray imaging procedures, as well as to the methods used to evaluate the quality of these procedures. Until recently, quality control procedures at health care institutions in Lithuania were limited to the testing of X-ray units. Since film processing is one of the most important factors influencing patient doses and image quality during X-ray examinations, in 2003 the Kaunas department of the Radiation Protection Centre organised inspections of film processing laboratories in 11 health care institutions-hospitals and outpatient departments-in the Kaunas region. Problems of non-compliance with requirements identified during these inspections are discussed in this paper. Most of the health care institutions inspected already had quality assurance programmes. However, the implementation of these programmes was sometimes erratic because of the insufficient attention paid to the film developing processes. The worst situation was found in 4 institutions where the films were developed manually. Only 3 of the 11 departments inspected had sensitometers and densitometers for quality control of the processing. In many cases there was no control of chemicals, film sensitivity and density, or else control was irregular. In only a few departments were the effects of repeated controls investigated and discussed. Despite the current problems occurring in medical X-ray diagnostic departments in Lithuania, the situation is rapidly improving. New equipment is being installed, new devices for quality control are being used and, last but not least, the view of hospital administrators, radiologists and laboratory workers towards quality assurance and quality control is changing.

摘要

鉴于诊断性X射线检查对图像质量要求很高,优化X射线成像以降低患者剂量是一个复杂的过程。当实施质量体系作为优化的基础时,应关注工作人员的资质、设备及X射线成像程序的质量控制,以及用于评估这些程序质量的方法。直到最近,立陶宛医疗机构的质量控制程序还仅限于对X射线设备的检测。由于胶片冲洗是影响X射线检查中患者剂量和图像质量的最重要因素之一,2003年,辐射防护中心考纳斯部门对考纳斯地区11家医疗机构(医院和门诊部)的胶片冲洗实验室进行了检查。本文讨论了这些检查中发现的不符合要求的问题。大多数接受检查的医疗机构已经有质量保证计划。然而,由于对胶片冲洗过程关注不足,这些计划的实施有时并不稳定。在4家手工冲洗胶片的机构中发现了最糟糕的情况。在接受检查的11个部门中,只有3个部门有用于冲洗质量控制的感光计和密度计。在许多情况下,没有对化学药品、胶片感光度和密度进行控制,或者控制不规范。只有少数部门对重复控制的效果进行了调查和讨论。尽管立陶宛医疗X射线诊断部门目前存在问题,但情况正在迅速改善。新设备正在安装,新的质量控制设备正在使用,最后但同样重要的是,医院管理人员、放射科医生和实验室工作人员对质量保证和质量控制的观念正在改变。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验