Abstract

A Monte Carlo computer model of mammography has been developed to study and optimise the performance of digital mammographic systems. The program uses high-resolution voxel phantoms to model the breast, which simulate the adipose and fibroglandular tissues, Cooper's ligaments, ducts and skin in three dimensions. The model calculates the dose to each tissue, and also the quantities such as energy imparted to image pixels, noise per image pixel and scatter-to-primary (S/P) ratios. It allows studies of the dependence of image properties on breast structure and on position within the image. The program has been calibrated by calculating and measuring the pixel values and noise for a digital mammographic system. The thicknesses of two components of this system were unknown, and were adjusted to obtain a good agreement between measurement and calculation. The utility of the program is demonstrated with the calculations of the variation of the S/P ratio with and without a grid, and of the image contrast across the image of a 50-mm-thick breast phantom.

摘要

已开发出一种用于乳腺钼靶摄影的蒙特卡罗计算机模型,以研究和优化数字乳腺钼靶系统的性能。该程序使用高分辨率体素模型来模拟乳房,在三维空间中模拟脂肪和纤维腺组织、库珀韧带、导管和皮肤。该模型计算每个组织的剂量,以及诸如赋予图像像素的能量、每个图像像素的噪声和散射与原发射线(S/P)比率等量值。它允许研究图像特性对乳房结构和图像内位置的依赖性。通过计算和测量数字乳腺钼靶系统的像素值和噪声,对该程序进行了校准。该系统的两个组件的厚度未知,通过调整使其在测量和计算之间取得良好的一致性。通过计算有无格栅时S/P比率的变化以及50毫米厚乳房模型图像上的图像对比度,证明了该程序的实用性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验