Threshold pixel size for shape determination of microcalcifications in digital mammography: a pilot study.

作者信息

Ruschin M, Hemdal B, Andersson I, Börjesson S, Håkansson M, Båth M, Grahn A, Tingberg A

机构信息

Department of Medical Radiation Physics, Lund University, Malmö University Hospital, Malmö, Sweden.

出版信息

Radiat Prot Dosimetry. 2005;114(1-3):415-23. doi: 10.1093/rpd/nch561.

Abstract

The effect of pixel size on shape determination in screening digital mammography systems was studied using a shape identification task as the measured outcome. Ten microcalcifications on screen-films were digitised to a range of pixel sizes (2.5-200 microm) and extracted from computed radiography (CR) images (50 microm) acquired under equivalent imaging conditions. Fifteen observers attempted to identify the shape of each microcalcification at each pixel size. The results were collated to provide a fraction of correct responses vs. pixel size curve for each microcalcification. Averaging over all shapes, pixel values >100 microm lead to a significant decrease in shape determination ability (p < 0.01) for digitised screen-film. For CR images, half the shapes were not properly identified. Hence, although 20-100 microm was sufficient for microcalcification shape determination for digitised screen-film images, 50 microm was only borderline sufficient for the CR digital images.