McKee Clayton T, Clark Spencer C, Walz John Y, Ducker William A
Department of Chemistry, Virginia Tech, Blacksburg, Virginia 24061, USA.
Langmuir. 2005 Jun 21;21(13):5783-9. doi: 10.1021/la046856p.
We describe measurements of the scattering of visible light from an evanescent field by both spherical particles (R = 1-10 mum) that are glued to atomic force microscopy (AFM) cantilevers, and by sharp tips (R < 60 nm) that were incorporated onto the cantilevers during manufacture. The evanescent wave was generated at the interface between a flat plate and an aqueous solution, and an atomic force microscope was used to accurately control the separation, h, between the particle and the flat plate. We find that, for sharp tips, the intensity of scattered light decays exponentially with separation between the tip and the plate all the way down to h approximately 0. The measured decay length of scattered intensity, delta, is the same as the theoretical decay length of the evanescent intensity in the absence of the sharp tip. For borosilicate particles, (R = 1-10 mum), the scattering also decays exponentially with separation at large separations. However, when the separation is less than roughly 3delta, the measured scattering intensity is smaller in magnitude than that which would be predicted by extrapolating the exponential decay observed at large separations. For these particles, the scattering approximately fits the sum of two exponentials. The magnitude of the deviation from exponential at contact was roughly 10-15% for R = 1 mum particles and about 30% for larger particles and is larger for s-polarized light. Preliminary experiments on polystyrene particles shows that the scattering is also smaller than exponential at small separations but that the deviation from exponential is larger for p-polarized light. In evanescent wave AFM (EW-AFM) the scattering-separation can be calibrated for situations where the scattering is not exponential. We discuss possible errors that could be introduced by assuming that exponential decay of scattering continues down to h = 0.
我们描述了对附着在原子力显微镜(AFM)悬臂上的球形颗粒(半径R = 1 - 10微米)以及在制造过程中集成到悬臂上的尖锐尖端(半径R < 60纳米)所产生的倏逝场中可见光散射的测量。倏逝波是在平板与水溶液的界面处产生的,并且使用原子力显微镜精确控制颗粒与平板之间的间距h。我们发现,对于尖锐尖端,散射光强度随着尖端与平板之间的间距呈指数衰减,直至h约为0。所测量的散射强度衰减长度δ与不存在尖锐尖端时倏逝强度的理论衰减长度相同。对于硼硅酸盐颗粒(半径R = 1 - 10微米),在大间距时散射也随间距呈指数衰减。然而,当间距小于约3δ时,所测量的散射强度大小比通过外推大间距时观察到的指数衰减所预测的值要小。对于这些颗粒,散射大致符合两个指数之和。对于半径R = 1微米的颗粒,接触时偏离指数的幅度约为10 - 15%,对于较大颗粒约为30%,并且对于s偏振光更大。对聚苯乙烯颗粒的初步实验表明,在小间距时散射也小于指数形式,但对于p偏振光,偏离指数的程度更大。在倏逝波原子力显微镜(EW - AFM)中,对于散射不是指数形式的情况,可以校准散射 - 间距关系。我们讨论了假设散射的指数衰减一直持续到h = 0可能引入的潜在误差。