Knirsch Walter, Dodge-Khatami Ali, Balmer Christian, Peuster Matthias, Kadner Alexander, Weiss Markus, Prêtre Rene, Berger Felix
Division of Pediatric Cardiology, University Children's Hospital, Zurich, Switzerland.
Catheter Cardiovasc Interv. 2005 Sep;66(1):10-3. doi: 10.1002/ccd.20436.
A 3-year-old boy underwent interventional closure of an atrial septal defect using an Amplatzer septal occluder. After 4 weeks, an aortic sinus-to-left atrial fistula was detected by echocardiography in an asymptomatic child. The device was surgically explanted with fistula and atrial septal defect closure. Follow-up was uneventful.