在拟人化结肠模型中进行电子减影清洁后CT结肠成像的息肉大小。

Polyp size at CT colonography after electronic subtraction cleansing in an anthropomorphic colon phantom.

作者信息

Zalis Michael E, Perumpillichira James J, Kim John Y, Del Frate Chiara, Magee Cordula, Hahn Peter F

机构信息

Division of Abdominal Imaging and Intervention, Department of Radiology, Massachusetts General Hospital, 55 Fruit St, White 270, Boston, MA 02114, USA.

出版信息

Radiology. 2005 Jul;236(1):118-24. doi: 10.1148/radiol.2361040231.

Abstract

PURPOSE

To evaluate the effect of various bowel contrast material concentrations and subtraction software on size measurements of well-defined polyp lesions in a colon phantom at CT colonography.

MATERIALS AND METHODS

Repeated scanning and a precise reference standard required the use of a colon phantom in which 21 polyps were randomly distributed. Two readers who had each reviewed computed tomographic (CT) colonographic images from more than 100 cases evaluated polyp size on images obtained when the phantom was partially filled with varying concentrations of contrast material, scanned by using CT colonography, and subjected to electronic subtraction cleansing. The single largest dimension was recorded for each reader for a randomized series of polyps. These measurements were compared with a reference standard that was based on a combination of the manufacturer's polyp size specifications and the subsequent verification of these sizes by an independent consensus panel. Six weeks after initial observations, readers evaluated images of the phantom scanned without the presence of contrast material. Polyp size estimations for the two readers for each series were compared with the reference standard to obtain a mean absolute measurement error for each reader for each series. Data for each reader were compared by using a nonparametric Kruskal-Wallis analysis of variance test. A pair-wise comparison of the experimental and control series was then performed by using the Dunn post hoc test.

RESULTS

Contrast material dilutions resulting in an average attenuation of less than 500 HU resulted in complete subtraction and the absence of streak artifacts. There was no statistically significant difference between the average measurement error for contrast attenuations between 300 and 500 HU when compared with that of control. Streak artifact was noticeable for the highest dilution (mean, 840 HU). No statistically significant differences were observed for series in which cleansing software was used in the absence of bowel contrast material.

CONCLUSION

The combination of electronic cleansing and bowel contrast enhancement in the range of 300-500 HU results in no substantial change in readers' estimations of polyp size at CT colonography.

摘要

目的

评估不同肠道造影剂浓度及减影软件对CT结肠成像中结肠模型内边界清晰的息肉病变大小测量的影响。

材料与方法

由于需要重复扫描及精确的参考标准,故使用了一个随机分布有21个息肉的结肠模型。两位各自审阅过100多例CT结肠成像图像的阅片者,对模型部分填充不同浓度造影剂、经CT结肠成像扫描并进行电子减影清洗后获得的图像上的息肉大小进行评估。对每个阅片者,记录一系列随机息肉中的单个最大尺寸。将这些测量结果与基于制造商息肉尺寸规格并经独立共识小组随后验证的参考标准进行比较。在初始观察六周后,阅片者评估未使用造影剂扫描的模型图像。将每个系列中两位阅片者的息肉大小估计值与参考标准进行比较,以获得每个阅片者每个系列的平均绝对测量误差。使用非参数Kruskal-Wallis方差分析检验对每个阅片者的数据进行比较。然后使用Dunn事后检验对实验系列和对照系列进行成对比较。

结果

造影剂稀释导致平均衰减小于500 HU时可实现完全减影且无条纹伪影。与对照组相比,300至500 HU之间的造影剂衰减平均测量误差无统计学显著差异。最高稀释度(平均840 HU)时条纹伪影明显。在未使用肠道造影剂而使用清洗软件的系列中未观察到统计学显著差异。

结论

在300 - 500 HU范围内,电子清洗与肠道造影增强相结合不会使阅片者对CT结肠成像中息肉大小的估计产生实质性变化。