Ghoshal Amit K, Guo Tiedong, Soukhova Nadia, Soldin Steven J
Department of Laboratory Medicine, Children's National Medical Center, 111 Michigan Avenue, NW, Washington, DC 20010-2970, USA.
Clin Chim Acta. 2005 Aug;358(1-2):104-12. doi: 10.1016/j.cccn.2005.02.011. Epub 2005 Apr 9.
Tandem mass spectrometry (MS/MS) is being increasingly used to identify and measure acylcarnitines in blood and urine of children suspected of having fatty oxidation disorders and other inborn errors of metabolism. Rapid MS/MS analysis requires simple and efficient sample preparation. We developed a LC-MS/MS method for the online extraction of acylcarnitines in plasma without derivatization that requires only precipitation of proteins by acetonitrile followed by centrifugation, thus increasing efficiency.
An API-3000 tandem mass spectrometer (SCIEX, Toronto, Canada) equipped with electrospray ionization (ESI), TurboIon Spray source, three Shimadzu LC10AD micropumps and autosampler (Shimadzu Scientific Instruments, Columbia, MD) was used to perform the analysis. Within-day and between-day imprecision was evaluated for 10 analytes in the MRM mode using 3 levels of controls. Accuracy was determined by comparing the method with another MS/MS procedure and by recovery experiments. Sensitivity and specificity were evaluated by identifying patient samples under a wide variety of clinical conditions.
Within-day CVs was <10% for all analytes tested and between-day CVs ranged from 4.4% to 14.2%. The method was linear in the range between 1.0 and 100 micromol/l for C2 and 0.1 and 10 micromol/l for the other acylcarnitines. The results of the comparison study yielded r values ranging between 0.948 and 0.999. Recovery ranged from 84% to 112%. The method correctly identified patients with a variety of fatty acid oxidation disorders and organic acidemias.
Our method is a simple procedure for the analysis of acylcarnitines in plasma with minimal sample preparation. It is thus ideal in a routine clinical setting where efficient processing of clinical samples is necessary to reduce turnaround time under conditions of high-throughput.
串联质谱法(MS/MS)越来越多地用于识别和测量疑似患有脂肪氧化障碍及其他先天性代谢缺陷的儿童血液和尿液中的酰基肉碱。快速MS/MS分析需要简单高效的样品制备方法。我们开发了一种无需衍生化的液相色谱-串联质谱法(LC-MS/MS),用于在线提取血浆中的酰基肉碱,该方法仅需用乙腈沉淀蛋白质,然后离心,从而提高了效率。
使用配备电喷雾电离(ESI)、TurboIon Spray源的API-3000串联质谱仪(SCIEX,加拿大多伦多),以及三个岛津LC10AD微量泵和自动进样器(岛津科学仪器公司,马里兰州哥伦比亚)进行分析。在多反应监测(MRM)模式下,使用3个浓度水平的质控品对10种分析物进行日内和日间精密度评估。通过将该方法与另一种MS/MS方法进行比较以及回收率实验来确定准确性。通过识别各种临床情况下的患者样本评估灵敏度和特异性。
所有测试分析物的日内变异系数(CV)均<10%,日间CV范围为4.4%至14.2%。该方法对于C2的线性范围为1.0至100 μmol/L,对于其他酰基肉碱为0.1至10 μmol/L。比较研究结果的r值范围为0.948至0.999。回收率范围为84%至112%。该方法正确识别了患有各种脂肪酸氧化障碍和有机酸血症的患者。
我们的方法是一种用于分析血浆中酰基肉碱的简单程序,样品制备最少。因此,在常规临床环境中,当需要高效处理临床样本以在高通量条件下缩短周转时间时,它是理想的方法。