Kahlert H, Pörksen J R, Behnert J, Scholz F
Institut für Chemie und Biochemie, Universität Greifswald, Soldmannstr. 23, 17489 Greifswald, Germany,
Anal Bioanal Chem. 2005 Aug;382(8):1981-6. doi: 10.1007/s00216-005-3379-x. Epub 2005 Jul 21.
A pH-sensitive detector for flow-through potentiometry based on a graphite/quinhydrone composite electrode was applied for flow-injection analysis (FIA) titrations. Hydrochloric acid and acetic acid were titrated by injection of samples into a sodium hydroxide carrier solution. System conditions were optimised by variation of flow rate, injection volume and titrant concentration. The parameters sampling frequency, residence time and dispersion coefficient were determined. The evaluation of peak width (time between the two inflection points on each side of the peak), peak area and slope of the raising edge of the peak lead were studied with respect to their use for calibration. Hydrochloric acid and acetic acid could be titrated down to a concentration of 2 x 10(-4) mol L(-1) using 150-microL injection volumes, which is almost ten times lower than can be achieved using colour indicators and a spectrophotometric detection.
一种基于石墨/醌氢醌复合电极的用于流通电位法的pH敏感检测器被应用于流动注射分析(FIA)滴定。通过将样品注入氢氧化钠载液中对盐酸和乙酸进行滴定。通过改变流速、进样体积和滴定剂浓度对系统条件进行了优化。测定了采样频率、停留时间和分散系数。研究了峰宽(峰两侧两个拐点之间的时间)、峰面积和峰前沿上升斜率用于校准的情况。使用150微升进样体积时,盐酸和乙酸可滴定至浓度低至2×10⁻⁴摩尔/升,这几乎比使用颜色指示剂和分光光度检测所能达到的浓度低十倍。