Epithermal neutron beams for clinical studies of boron neutron capture therapy: a dosimetric comparison of seven beams.

作者信息

Binns P J, Riley K J, Harling O K

机构信息

Nuclear Reactor Laboratory, Massachusetts Institute of Technology, 138 Albany Street, Cambridge, Massachusetts 02139, USA.

出版信息

Radiat Res. 2005 Aug;164(2):212-20. doi: 10.1667/rr3404.

Abstract

A comparison of seven epithermal neutron beams used in clinical studies of boron neutron capture therapy (BNCT) in Sweden (Studsvik), Finland (Espoo), Czech Republic (ReZ), The Netherlands (Petten) and the U.S. (Brookhaven and Cambridge) was performed to facilitate sharing of preclinical and clinical results. The physical performance of each beam was measured using a common dosimetry method under conditions pertinent to brain irradiations. Neutron fluence and absorbed dose measurements were performed with activation foils and paired ionization chambers on the central axis both in air and in an ellipsoidal water phantom. The overall quality of each beam was assessed by figures of merit determined from the total weighted dose profiles that assumed the presence of boron in tissue. The in-air specific beam contamination from both fast neutrons and gamma rays ranged between 8 and 65 x 10(-11) cGy(w) cm2 for the different beams and the epithermal neutron flux intensities available at the patient position differed by more than a factor of 20 from 0.2-4.3 x 10(9) n cm(-2) s(-1). Percentage depth dose profiles measured in-phantom for the individual photon, thermal and fast-neutron dose components differed only subtly in shape between facilities. Assuming uptake characteristics consistent with the currently used boronated phenylalanine, all the epithermal beams exhibit a useful penetration of 8 cm or greater that is sufficient to irradiate a lesion seated at the brain midline. The performance of the existing facilities will benefit from the introduction of advanced compounds through improved beam penetrability. This could increase by as much as 2 cm for the purest of beams, although the beam intensities generally need to be increased to between 2-5 x 10(9) n cm(-2) s(-1) to maintain manageable irradiation times. These data provide the first consistent measurement of beam performance at the different centers and will enable a preliminary normalization of the calculated patient dosimetry.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验