Acta Neurochir (Wien). 1992;115(3-4):152-5. doi: 10.1007/BF01406375.
The use of a 20 MHz pulsed Doppler technique during stereotactic brain tumour biopsies is described. By means of a miniaturized ultrasonic probe the tissue planned for biopsy was investigated for vessel signals. Out of 41 patients, with this technique a flow-pulse curve in front of the biopsy cannula was registered in 14 cases. In 9 cases one pulse curve was determined and in 5 cases two or more flow curves were found, indicating the presence of arterial vessels. The depth of the vessels signal could be localized and this area avoided for biopsy sampling. In a series of 41 stereotactic biopsies carried out in this way there was no intra-operative haemorrhage or postoperative haematoma.
本文描述了在立体定向脑肿瘤活检过程中使用20兆赫脉冲多普勒技术的情况。通过一个小型化超声探头,对计划活检的组织进行血管信号检测。在41例患者中,采用该技术在14例患者的活检套管前方记录到了血流脉冲曲线。其中9例确定为一条脉冲曲线,5例发现两条或更多血流曲线,提示存在动脉血管。血管信号的深度能够被定位,活检采样时可避开该区域。以这种方式进行的一系列41例立体定向活检中,未发生术中出血或术后血肿。