Brüchert Wolfram, Bettmer Jörg
Institute of Inorganic Chemistry and Analytical Chemistry, University of Mainz, Duesbergweg 10-14, D-55099 Mainz, Germany.
Anal Chem. 2005 Aug 1;77(15):5072-5. doi: 10.1021/ac050425+.
The on-line coupling of gel electrophoresis (GE) and inductively coupled plasma-mass spectrometry (ICPMS) is described for the first time. The new method combines the separation power of GE for large biomolecules and the high sensitivity and elemental selectivity of ICPMS. This coupling has been achieved by means of gels housed in glass tubes (length 2.5-20 cm; i.d. 0.5-5 mm). The gel is fixed by glass frits permeable for the analytes. After the electrophoretic separation at voltages up to 500 V, the analytes are transferred to the nebulizer of the ICPMS by an eluent stream, which is separated from the electrode chamber by a membrane. This filter blocks molecules with molecular masses larger than 500 Da. Using double-stranded DNA fragments, commercially available standard solutions were analyzed with on-line GE-ICPMS for the first time by monitoring (31)P(+) with a double-focusing mass spectrometer at a mass resolution of 4000.
首次描述了凝胶电泳(GE)与电感耦合等离子体质谱(ICPMS)的在线联用。新方法结合了GE对大分子的分离能力以及ICPMS的高灵敏度和元素选择性。这种联用是通过置于玻璃管(长度2.5 - 20 cm;内径0.5 - 5 mm)中的凝胶实现的。凝胶由对分析物可渗透的玻璃料固定。在高达500 V的电压下进行电泳分离后,分析物通过洗脱液流转移至ICPMS的雾化器,该洗脱液流通过膜与电极室分离。此过滤器可阻挡分子量大于500 Da的分子。使用双链DNA片段,首次通过在线GE - ICPMS,在质量分辨率为4000的情况下,用双聚焦质谱仪监测(31)P(+),分析了市售标准溶液。