微球法测定心输出量的评估

Evaluation of the microsphere-method for determination of cardiac output.

作者信息

Segadal L, Svanes K

出版信息

Scand J Clin Lab Invest. 1979 Sep;39(5):415-21. doi: 10.3109/00365517909106126.

Abstract

Cardiac output was determined by means of radioactive microspheres, 15 +/- 5 microns in diameter. Blood flow in the ascending aorta was measured by an electromagnetic flowmeter. In eight cats thirty-eight simultaneous measurements were made of cardiac output and aortic flow. The observations correlated well (r = 0.89), with a mean difference of 5.2%, probably corresponding to coronary blood flow. Continuous flow recordings showed no alterations in aortic flow during the injections of microspheres; arterial blood pressure and heart rate remained unchanged even after six injections, each consisting of about 2 x 10(5) spheres per kg body weight. Tests for shunting of microspheres were performed and revealed shunt fractions in the systemic circulation of about 8%, with no significant shunting through the lungs. The microsphere method for determining cardiac output is thought to be an accurate method, suitable for small and medium sized animals.

摘要

采用放射性微球(直径15±5微米)测定心输出量。用电磁流量计测量升主动脉血流。对8只猫同时进行了38次心输出量和主动脉血流测量。观察结果相关性良好(r = 0.89),平均差异为5.2%,可能相当于冠状动脉血流量。连续血流记录显示,注射微球期间主动脉血流无变化;即使在每次注射约2×10⁵个/千克体重微球,共注射6次后,动脉血压和心率仍保持不变。进行了微球分流测试,结果显示体循环中的分流分数约为8%,肺部无明显分流。测定心输出量的微球法被认为是一种准确的方法,适用于中小型动物。