Ricalde Pat, Engroff Stephen L, Von Fraunhofer J Anthony, Posnick Jeffrey C
University of Maryland Dental School, Baltimore, MD, USA.
J Oral Maxillofac Surg. 2005 Aug;63(8):1180-3. doi: 10.1016/j.joms.2005.04.019.
Mandibulotomy is used to access various tumors of the tongue base, posterior oral cavity, pharynx, parapharyngeal space, and cranial base. Internal fixation using titanium plates and screws is the most common method of stabilization. These have the potential for interference with radiotherapy delivery. This in vitro study compares the strength of titanium and resorbable internal fixation in a mandibulotomy model by analyzing the force required for plate and screw breakage.
Red oak wood board was used to simulate the mandible. Titanium and resorbable plates and screws in various configurations were used to stabilize pieces of the wood. They were arranged in 6 different groups. The specimens were individually tested with a vertical load, while the test machine recorded the force-versus-displacement behavior automatically.
Plate type and configuration affected the applied load required to induce displacement of the simulated mandibulotomy. Heating and cooling the resorbable plates prior to strength testing also affected the load-versus-displacement curve.
Overall, the titanium system we studied exhibited greater resistance to deformation from a vertical load than did the resorbable plate groups.
下颌骨切开术用于治疗舌根、口腔后部、咽部、咽旁间隙和颅底的各种肿瘤。使用钛板和螺钉进行内固定是最常见的稳定方法。这些可能会干扰放射治疗的实施。这项体外研究通过分析钛板和螺钉断裂所需的力,比较了钛和可吸收内固定在下颌骨切开术模型中的强度。
使用红橡木木板模拟下颌骨。使用各种配置的钛板和可吸收板及螺钉来固定木块。它们被分为6个不同的组。对标本分别施加垂直载荷进行测试,测试机自动记录力与位移的关系。
板的类型和配置影响模拟下颌骨切开术位移所需的外加负荷。在强度测试前对可吸收板进行加热和冷却也会影响负荷与位移曲线。
总体而言,我们研究的钛系统比可吸收板组对垂直载荷引起的变形具有更大的抵抗力。