Olsrud Johan, Lätt Jimmy, Brockstedt Sara, Romner Bertil, Björkman-Burtscher Isabella M
MR Department, Lund University Hospital, Lund, Sweden.
J Magn Reson Imaging. 2005 Sep;22(3):433-7. doi: 10.1002/jmri.20391.
To evaluate artifact sizes at 3 T compared to at 1.5 T, and to evaluate the influence of scanning parameters with respect to artifact size on a 3-T magnetic resonance imaging (MRI) system.
Two aneurysm clips and five shunt valves were imaged in a water phantom at 1.5 and 3 T. At 3 T the influence of bandwidth (spin echo (SE) images) and echo time (gradient echo (GRE) images) on artifact size (area and extension in two orthogonal directions) was investigated.
Artifact sizes increased substantially (typically 5-10 mm) at 3 T, compared to at 1.5 T, for implants entirely made of metallic materials, whereas the increase was the size less prominent (0-5 mm) for implants only partly containing metal. Artifact areas could be altered by changing the bandwidth or the echo time to about the same extent as it was affected by the increased field strength.
Artifact sizes increase at 3 T, compared to at 1.5 T, depending on the type and composition of the implant, but can be substantially reduced by altering the imaging parameters. Optimization of imaging protocols to minimize artifacts is therefore important at higher field strengths.
评估3 T时与1.5 T时伪影的大小,并评估扫描参数对3 T磁共振成像(MRI)系统上伪影大小的影响。
在水模中对两个动脉瘤夹和五个分流阀在1.5 T和3 T下进行成像。在3 T时,研究带宽(自旋回波(SE)图像)和回波时间(梯度回波(GRE)图像)对伪影大小(在两个正交方向上的面积和范围)的影响。
对于完全由金属材料制成的植入物,与1.5 T时相比,3 T时伪影大小显著增加(通常为5 - 10 mm);而对于仅部分含有金属的植入物,增加幅度较小(0 - 5 mm)。通过改变带宽或回波时间,伪影面积的改变程度与受场强增加的影响程度大致相同。
与1.5 T时相比,3 T时伪影大小会增加,这取决于植入物的类型和成分,但可以通过改变成像参数大幅减小。因此,在更高场强下优化成像方案以最小化伪影很重要。