Suppr超能文献

人工腰椎间盘的磁共振成像:安全性与金属伪影

Magnetic resonance imaging of artificial lumbar disks: safety and metal artifacts.

作者信息

Yang Chang-wei, Liu Liu, Wang Jian, Dong Ai-sheng, Lu Jian-ping, He Shi-sheng, Li Ming

机构信息

Orthopedics Department, Changhai Hospital, Second Military Medical University, Shanghai 200433, China.

出版信息

Chin Med J (Engl). 2009 Apr 20;122(8):911-6.

Abstract

BACKGROUND

This study was to investigate the safety of two types of commercially available lumbar artificial discs (CHARITE and PRODISC-L) during a magnetic resonance imaging (MRI) procedure in a 1.5-Tesla MR system, and to evaluate the size of metal artifacts on the MR image for different sequences.

METHODS

A 1.5-Tesla clinical MR imaging system was used. The degree of deflection of the endplates of two artificial discs was evaluated by an angle-measurement instrument at the portals of the MRI scanner. The heating effect of the radio frequency (RF) magnetic field was evaluated by using "worst-case" imaging sequences on a human cadaver implanted with an artificial lumbar disc at the L5/S1 intervertebral disc location. The temperatures of the tissue adjacent to the implant, and of the L4/L5 intervertebral disc (used as a control) were measured, respectively, using a digital probe thermometer before and after the MRI scan sequence. A rectangular water phantom was designed to evaluate the metal artifacts of these two artificial discs under different MR imaging sequences.

RESULTS

The maximal deflection angle of the endplate of the implants under a static MR field was 7.5 and 6.0 degrees, for the CHARITE and PRODISC-L, respectively. The difference between temperature rise of tissue adjacent to the two types of artificial discs and the temperature rise of the L4/L5 control location was 0.4 and 0.6 degrees C, respectively. The size of metal artifacts on images of TSE (T1/T2 -weighted), STIR and Turbo Dark Fluid sequences were relatively less than those of TSE fat saturation, Flash and SE (T1-weighted) sequences.

CONCLUSIONS

The CHARITE and the PRODISC-L artificial disc do not present an additional hazard or risk to a patient undergoing an MRI procedure using a scanner operating with a static magnetic field of 1.5 T or lower. Image artifacts from the implants may present problems if the anatomical region of interest is in or near the area where these implants are located (e.g., vertebral canal at affected segment).

摘要

背景

本研究旨在调查两种市售腰椎人工椎间盘(CHARITE和PRODISC-L)在1.5特斯拉磁共振成像(MRI)系统的MRI检查过程中的安全性,并评估不同序列MR图像上金属伪影的大小。

方法

使用1.5特斯拉临床MR成像系统。通过MRI扫描仪入口处的角度测量仪评估两个人工椎间盘终板的偏斜程度。在L5/S1椎间盘位置植入人工腰椎间盘的人体尸体上,使用“最坏情况”成像序列评估射频(RF)磁场的热效应。在MRI扫描序列前后,分别使用数字探头温度计测量植入物附近组织以及L4/L5椎间盘(用作对照)的温度。设计一个矩形水模体来评估这两个人工椎间盘在不同MR成像序列下的金属伪影。

结果

在静态MR场下,CHARITE和PRODISC-L植入物终板的最大偏斜角度分别为7.5度和6.0度。两种人工椎间盘附近组织的温度升高与L4/L5对照位置的温度升高之间的差异分别为0.4摄氏度和0.6摄氏度。TSE(T1/T2加权)、STIR和Turbo Dark Fluid序列图像上的金属伪影大小相对小于TSE脂肪饱和、Flash和SE(T1加权)序列。

结论

CHARITE和PRODISC-L人工椎间盘对于在1.5 T或更低静态磁场的扫描仪上进行MRI检查的患者不会带来额外的危害或风险。如果感兴趣的解剖区域位于这些植入物所在区域或其附近(例如,受影响节段的椎管),植入物产生的图像伪影可能会带来问题。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验