The structural principles and technical performances of different ionization chamber dosimeters used in radiotherapy have been briefly introduced in this paper. In the mean time, a comparison of technical performances of four different kinds of ionization chamber dosimeter used in hospitals in China has also been made. The data show their quality meet the required specification specified by IEC731 and JJG912-96, so they can be used as field survey dosimeters in hospitals. Some precautions in using these dosimeters are given here too.

摘要

本文简要介绍了放射治疗中使用的不同电离室剂量仪的结构原理和技术性能。同时,还对中国医院使用的四种不同类型电离室剂量仪的技术性能进行了比较。数据表明,它们的质量符合国际电工委员会(IEC)731标准和中华人民共和国国家计量检定规程JJG912 - 96的要求,因此可作为医院的现场测量剂量仪。文中还给出了使用这些剂量仪时的一些注意事项。