Li Zheng-ping, Li Yan-kun, Wang Yu-cong
College of Chemistry and Environment Science, Hebei University, Baoding 071002, Hebei Province, People's Republic of China.
Luminescence. 2005 Jul-Oct;20(4-5):282-6. doi: 10.1002/bio.835.
The effect of hexa-amine cobalt cations on the DNA condensation in aqueous solution was investigated by resonance light scattering (RLS). When the relative concentration of hexa-amine cobalt (III) cations to DNA is in the appropriate range, the cations will induce DNA condensation and aggregation, which results in a strong RLS spectrum characterized by a peak at 290.0 nm. The RLS technique is a powerful tool for monitoring DNA condensation and, under optimal conditions, the enhanced RLS intensity at 290.0 nm was proportional to the concentration of DNA in the range 0.01-6.0 microg/mL. Based on this, a sensitive and convenient analysis method for the microdetermination of DNA was established. The detection limit (3 s) of calf thymus DNA by the proposed method is 1.9 ng/mL and few substances interfere in the DNA determination.
采用共振光散射(RLS)技术研究了六胺钴阳离子对水溶液中DNA凝聚的影响。当六胺钴(III)阳离子与DNA的相对浓度处于适当范围内时,阳离子会诱导DNA凝聚和聚集,从而产生以290.0 nm处的峰为特征的强烈共振光散射光谱。共振光散射技术是监测DNA凝聚的有力工具,在最佳条件下,290.0 nm处增强的共振光散射强度与0.01 - 6.0 μg/mL范围内的DNA浓度成正比。基于此,建立了一种灵敏、简便的微量测定DNA的分析方法。该方法测定小牛胸腺DNA的检测限(3s)为1.9 ng/mL,很少有物质干扰DNA的测定。