Kistler Peter M, Kalman Jonathan M, Fynn Simon P, Singarayar Suresh, Roberts-Thomson Kurt C, Lindsay Catherine B, Khong Uyen, Sparks Paul B, Strathmore Neil, Mond Harry G
Department of Cardiology, The Royal Melbourne Hospital, Victoria 3050, Australia.
Pacing Clin Electrophysiol. 2005 Sep;28(9):903-9. doi: 10.1111/j.1540-8159.2005.00209.x.
There is an increasing use of active-fixation leads for cardiac pacing, yet concerns remain regarding initial high stimulation thresholds. The aim was to perform a detailed analysis of pacing parameters at the time of implantation to determine when lead repositioning should be considered.
We performed a prospective observational study of consecutive new pacemaker implants. Detailed analysis of pacing parameters was collected at 2-minute intervals for 10 minutes, and at day 1 and week 8 following implant.
Ninety-four patients underwent implantation of 79 dual-chamber and 15 single-chamber pacemakers using active-fixation leads in both chambers. An initial threshold of >1 V was demonstrated in 45/94 (48%) ventricular leads (mean threshold 1.5 +/- 0.3 V). This declined rapidly to 0.9 +/- 0.3 V at 4 minutes (P < 0.01), 0.7 +/- 0.3 V at 10 minutes (P < 0.01), and 0.6 +/- 0.3 V at day 1 (P < 0.01). At day 1, 43/45 leads were <1 V. There were 79 atrial leads. An initial threshold of >1 V (mean 1.7 +/- 0.6 V) was demonstrated in 41/79 (52%) leads falling significantly to 1.1 +/- 0.5 V at 4 minutes (P < 0.01), 0.9 +/- 0.4 V at 10 minutes (P < 0.01), and 0.6 +/- 0.2 V at day 1 (P < 0.01). At 10 minutes, 32 of 41 leads demonstrated a threshold of <1 V with all leads <1 V at day 1. Thresholds were maintained medium term.
Active-fixation leads are commonly associated with initially high thresholds that fall rapidly. An initial threshold of 2 V should be provisionally accepted and retested at 4 minutes. The majority will have a threshold of <1 V the following day. A failure of a high threshold to decline at 4 minutes requires lead repositioning.
主动固定电极导线在心脏起搏中的应用日益增加,但对于其初始高刺激阈值仍存在担忧。本研究旨在对植入时的起搏参数进行详细分析,以确定何时应考虑重新定位电极导线。
我们对连续的新起搏器植入患者进行了一项前瞻性观察研究。在植入后的10分钟内,每隔2分钟收集一次起搏参数的详细分析数据,并在植入后的第1天和第8周进行收集。
94例患者植入了79台双腔起搏器和15台单腔起搏器,两个腔室均使用主动固定电极导线。45/94(48%)的心室电极导线初始阈值>1V(平均阈值1.5±0.3V)。该阈值在4分钟时迅速降至0.9±0.3V(P<0.01),10分钟时降至0.7±0.3V(P<0.01),第1天时降至0.6±0.3V(P<0.01)。在第1天时,43/45的电极导线阈值<1V。共有79根心房电极导线。41/79(52%)的电极导线初始阈值>1V(平均1.7±0.6V),在4分钟时显著降至1.1±0.5V(P<0.01),10分钟时降至0.9±0.4V(P<0.01),第1天时降至0.6±0.2V(P<0.01)。在10分钟时,41根电极导线中有32根的阈值<1V,在第1天时所有电极导线的阈值均<1V。阈值在中期保持稳定。
主动固定电极导线通常与初始高阈值相关,且该阈值会迅速下降。初始阈值为2V时应暂时接受,并在4分钟时重新测试。大多数电极导线在第二天的阈值将<1V。如果高阈值在4分钟时未下降,则需要重新定位电极导线。