Gentry Kenneth L, Sachedina Nasheer, Smith Stephen W
Department of Biomedical Engineering, Duke University, Durham, NC 27708, USA.
Ultrason Imaging. 2005 Apr;27(2):89-100. doi: 10.1177/016173460502700203.
The feasibility of catheter single-element ultrasound transducers for cardiac ablation has been shown previously. We describe the design and testing of catheter-sized linear phased arrays transducers for ultrasound ablation. One array has 86 PZT-4 elements operating at 8 MHz and 5 MHz. The overall array size is 14.9 mm by 3.1 mm (10 Fr). The other array has 50 PZT-5 elements operating at 4 MHz and is 17 mm by 3.1 mm (10 Fr). In order to produce the intensity needed to create lesions in heart tissue, we modified a real-time, 3D scanner to produce 100 Vpp 256-cycle transmit pulses at a pulse repetition frequency of 14.1 kHz. This made it possible for the PZT-4 and PZT-5 transducers to produce ISPTA of 3.26 W/cm2 and 142 W/cm2, respectively. When driving the transducers at high duty factor, the transmit circuitry in the scanner was damaged. A mechanically-focused transducer with the same dimensions as the PZT-4 transducer was built. When transmitting continuously at 9 MHz, it produced an ISPTA of 29.3 W/cm2. This created a lesion 5 mm across and 5 mm deep in beef tissue while raising the focal temperature 23 degrees C. Ablation is within the capabilities of a catheter phased array transducer integrated into a diagnostic ultrasound scanner.
导管单元素超声换能器用于心脏消融的可行性此前已得到证实。我们描述了用于超声消融的导管尺寸线性相控阵换能器的设计与测试。一个阵列有86个PZT - 4元件,工作频率为8兆赫和5兆赫。阵列整体尺寸为14.9毫米×3.1毫米(10F)。另一个阵列有50个PZT - 5元件,工作频率为4兆赫,尺寸为17毫米×3.1毫米(10F)。为了产生在心脏组织中形成损伤所需的强度,我们对一台实时三维扫描仪进行了改装,使其能以14.1千赫的脉冲重复频率产生100伏峰 - 峰值、256周期的发射脉冲。这使得PZT - 4和PZT - 5换能器分别能够产生3.26瓦/平方厘米和142瓦/平方厘米的空间峰值时间平均声强(ISPTA)。当以高占空比驱动换能器时,扫描仪中的发射电路受损。制造了一个尺寸与PZT - 4换能器相同的机械聚焦换能器。当以9兆赫连续发射时,它产生的ISPTA为29.3瓦/平方厘米。这在牛肉组织中形成了一个直径5毫米、深5毫米的损伤,同时使焦点温度升高了23摄氏度。消融在集成于诊断超声扫描仪中的导管相控阵换能器的能力范围内。