Baillif S, Chiquet C, Werner L, Burillon C, Denis P
Service d'0phtalmologie, Centre Hospitalier Universitaire Edouard Herriot, Lyon, France.
J Fr Ophtalmol. 2005 Oct;28(8):e4. doi: 10.1016/s0181-5512(05)81009-5.
To report pathological and histochemical analysis of an explanted opacified hydrophilic acrylic intraocular lens.
and method: A hydrophilic acrylic intraocular lens--Orion IFP3D6--was explanted 2 years after routine phacoemulsification because of its opacification associated with significant visual impairment (20/40). The intraocular lens was examined by gross and light microscopy. Full-thickness cut sections of the lens were stained with 1% alizarin red (a special stain for calcium).
Microscopic analyses revealed the presence of granules of various sizes distributed in a line parallel to the anterior and posterior curvatures of the lens. The granules stained positively for alizarin red.
Such opacification has been described with the SC60B-OUV, MDR Inc., which is also a one-piece foldable-design intraocular lens (28% hydrophilic material). The opacification of the intraocular lens occurred over various periods after surgery. Forty percent of these lenses have been explanted. Microscopic and histochemical evaluation of these intraocular lenses revealed the same type of granulation deposits within the lenses. The origin and mechanism of the opacification are still not fully understood.
报告一枚取出的混浊亲水性丙烯酸酯人工晶状体的病理及组织化学分析结果。
一枚亲水性丙烯酸酯人工晶状体——Orion IFP3D6,在常规超声乳化白内障吸除术后2年因混浊且伴有明显视力损害(20/40)而被取出。对该人工晶状体进行大体及光学显微镜检查。晶状体的全层切片用1%茜素红(一种钙的特殊染色剂)染色。
显微镜分析显示存在各种大小的颗粒,这些颗粒沿平行于晶状体前后曲率的方向呈线状分布。这些颗粒茜素红染色呈阳性。
MDR公司的SC60B - OUV人工晶状体也出现过这种混浊情况,它也是一体式可折叠设计的人工晶状体(含28%亲水性材料)。人工晶状体混浊在术后不同时间段出现。其中40%的此类晶状体已被取出。对这些人工晶状体的显微镜及组织化学评估显示晶状体内部有相同类型的颗粒状沉积物。混浊的起源和机制仍未完全明确。