Towards a standardization of thrombin generation assessment: the influence of tissue factor, platelets and phospholipids concentration on the normal values of Thrombogram-Thrombinoscope assay.

作者信息

Gerotziafas Grigoris T, Depasse François, Busson Joël, Leflem Lena, Elalamy Ismail, Samama Meyer M

机构信息

Service d'Hématologie Biologique, Hôpital Hôtel-Dieu de Paris, France.

出版信息

Thromb J. 2005 Oct 26;3:16. doi: 10.1186/1477-9560-3-16.

Abstract

BACKGROUND

Thrombin generation assay was developed several years ago to mimic physiological coagulation mechanisms but it had important limitations. Thrombogram-Thrombinoscope assay using a fluorogenic substrate, allows obtaining thrombin generation curves in non-defibrinated platelet rich plasma (PRP) in a fully automated manner.

METHODS

We standardised the methodology of Thrombogram-Thrombinoscope and we evaluated the precision of thrombin generation parameters (lag-time, maximum concentration of thrombin [Cmax], time required to reach Cmax [Tmax] and endogenous thrombin potential ETP) using different concentrations of recombinant human tissue factor, platelets or phospholipids. Normal values of thrombin generation assay were established in optimal experimental conditions.

RESULTS

In the presence of low TF concentrations (final dilution of thromboplastin in plasma: 1/1000-1/2000) the Thrombogram assay showed intra-assay and inter-assay coefficients of variation lower than 9%. Thrombin generation parameters showed an important inter-individual variability and the coefficients of variation ranged from 18% to 50%. In PRP the lag-time, Cmax and Tmax but not the ETP, were influenced by TF concentration. Thrombin generation parameters were not influenced by variations of platelet concentration from 50 x 10(9)/l to 400 x 10(9)/l. The addition of synthetic procoagulant phospholipids in PPP strongly influenced all the parameters of thrombogram. For all the parameters of thrombogram a threshold effect was observed in the presence of phospholipid concentrations equal or higher to 4 microM. In frozen-thawed PRP the lag-time and the Tmax were significantly reduced and the Cmax was increased compared to the fresh PRP, but the ETP, the intra assay and the inter-assay coefficients of variation were similar in both test-systems.

CONCLUSION

Thrombogram-Thrombinoscope assay performed in fresh or in frozen-thawed PRP has an acceptable precision, with low inter-assay and intra-assay coefficient of variations. The concentration of TF is determinant for the normal values of the studied parameters of thrombin generation. When the assay is performed in PPP, thrombin generation parameters are influenced by the concentration of procoagulant synthetic phospholipids. The optimal experimental conditions were obtained in the presence of 1/1000 final dilution of thromboplastin, a platelet count higher than 50 x 10(9)/l and a synthetic phospholipid concentration higher than 4 microM.

摘要

背景

凝血酶生成检测法是数年前开发出来用于模拟生理凝血机制的,但它存在重要局限性。使用荧光底物的血栓图 - 凝血酶监测仪检测法能够以全自动方式在未脱纤维蛋白的富血小板血浆(PRP)中获得凝血酶生成曲线。

方法

我们对血栓图 - 凝血酶监测仪的方法进行了标准化,并使用不同浓度的重组人组织因子、血小板或磷脂评估了凝血酶生成参数(延迟时间、凝血酶最大浓度 [Cmax]、达到 Cmax 所需时间 [Tmax] 和内源性凝血酶潜力 ETP)的精密度。在最佳实验条件下确定了凝血酶生成检测的正常值。

结果

在低组织因子(TF)浓度存在时(血浆中凝血活酶的最终稀释度:1/1000 - 1/2000),血栓图检测显示批内和批间变异系数低于 9%。凝血酶生成参数显示出个体间的重要变异性,变异系数范围为 18%至 50%。在 PRP 中,延迟时间、Cmax 和 Tmax 受 TF 浓度影响,但 ETP 不受影响。血小板浓度从 50×10⁹/L 变化至 400×10⁹/L 时,凝血酶生成参数不受影响。在 PPP 中添加合成促凝磷脂强烈影响血栓图的所有参数。对于血栓图的所有参数,在磷脂浓度等于或高于 4μM 时观察到阈值效应。与新鲜 PRP 相比,冻融 PRP 中的延迟时间和 Tmax 显著缩短,Cmax 增加,但两种检测系统中的 ETP、批内和批间变异系数相似。

结论

在新鲜或冻融 PRP 中进行的血栓图 - 凝血酶监测仪检测具有可接受的精密度,批内和批间变异系数较低。TF 浓度是所研究的凝血酶生成参数正常值的决定因素。当在 PPP 中进行检测时,凝血酶生成参数受促凝合成磷脂浓度影响。在凝血活酶最终稀释度为 1/1000、血小板计数高于 50×10⁹/L 和合成磷脂浓度高于 4μM 的情况下可获得最佳实验条件。