Effect of extraluminal oxygen on carbon dioxide laser ignition of endotracheal tubes.

作者信息

Arnold J E, Allphin A L

机构信息

Department of Otolaryngology-Head and Neck Surgery, Case Western Reserve University School of Medicine, Rainbow Babies and Children's Hospital, Cleveland, Ohio 44106.

出版信息

Arch Otolaryngol Head Neck Surg. 1992 Jul;118(7):722-4. doi: 10.1001/archotol.1992.01880070052010.

Abstract

Ignition times of endotracheal tubes used in laser surgery have been determined in room air with oxygen flowing through the tube. Using uncuffed endotracheal tubes in children, we measured the oxygen level around the endotracheal tube at the vocal cords and found that it closely approximates the inspired oxygen concentration. We then tested Rusch red rubber and polyvinyl chloride endotracheal tubes while varying the external oxygen concentration. Little difference in ignition time was found for the polyvinyl chloride tube. The red rubber tube ignited significantly faster as the ambient oxygen levels were increased to as little as 30%. Our tests indicate that uncuffed red rubber tubes are not safer than polyvinyl chloride tubes when used with the carbon dioxide laser. All endotracheal tubes used with the laser should be tested for ignition at various extraluminal oxygen concentrations.

摘要

已在室内空气中,在有氧气流经气管内导管的情况下测定了用于激光手术的气管内导管的点燃时间。在儿童中使用无套囊气管内导管时,我们测量了声带处气管内导管周围的氧气水平,发现其与吸入氧浓度非常接近。然后,我们在改变外部氧气浓度的同时测试了Rusch红色橡胶和聚氯乙烯气管内导管。聚氯乙烯导管的点燃时间几乎没有差异。随着环境氧气水平增加至仅30%,红色橡胶导管显著更快被点燃。我们的测试表明,使用二氧化碳激光时,无套囊红色橡胶导管并不比聚氯乙烯导管更安全。所有与激光一起使用的气管内导管都应在各种管外氧气浓度下进行点燃测试。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验