Kohler T, Beyeler M, Webster D, Müller R
Institute for Biomedical Engineering, Swiss Federal Institute of Technology (ETH) and University of Zürich, Moussonstrasse 18, 8044 Zürich, Switzerland.
Calcif Tissue Int. 2005 Nov;77(5):281-90. doi: 10.1007/s00223-005-0039-2. Epub 2005 Nov 4.
Microcomputed tomography (microCT) is widely used for nondestructive bone phenotyping in small animals, especially in the mouse. Here, we investigated the reproducibility and resolution dependence of microCT analysis of microstructural parameters in three different compartments in the mouse femur. Reproducibility was assessed with respect to precision error (PE%CV) and intraclass correlation coefficient (ICC). We examined 14 left femurs isolated postmortem from two strains of mice (seven per group). Measurements and analyses were repeated five times on different days. In a second step, analysis was repeated again five times for a single measurement. Resolution dependence was assessed by high-resolution measurements (10 microm) in one strain and subsequent image degrading. Reproducibility was better in full bone compartment and in cortical bone compartment in the diaphysis (PE%CV = 0.06-2.16%) than in trabecular compartment in the distal metaphysis (PE(%CV) = 0.59-5.24%). Nevertheless, ICC (0.92-1.00) showed a very high reliability of the assessed parameters in all regions, indicating very small variances within repeated measurements compared to the population variances. Morphometric indices computed from lower- and higher-resolution images displayed in general only weak dependence and were highly correlated with each other (R2 = 0.91-0.99). The results show that parameters in the full and cortical compartments were very reproducible, whereas precision in the trabecular compartment was somewhat lower. Nevertheless, all compartmental analysis methods were very robust, as shown by the high ICC values, demonstrating high suitability for application in inbred strains, where highest precision is needed due to small population variances.
微计算机断层扫描(microCT)广泛用于小动物的非破坏性骨表型分析,尤其是在小鼠中。在此,我们研究了小鼠股骨三个不同区域微结构参数的微CT分析的可重复性和分辨率依赖性。通过精度误差(PE%CV)和组内相关系数(ICC)评估可重复性。我们检查了从两种品系小鼠死后分离的14根左股骨(每组7根)。在不同日期重复测量和分析5次。第二步,对单次测量再次重复分析5次。通过对一个品系进行高分辨率测量(10微米)并随后进行图像降质来评估分辨率依赖性。骨干的全骨区域和皮质骨区域的可重复性(PE%CV = 0.06 - 2.16%)优于远端干骺端的小梁区域(PE(%CV) = 0.59 - 5.24%)。然而,ICC(0.92 - 1.00)表明所有区域评估参数的可靠性非常高,表明与总体方差相比,重复测量中的方差非常小。从低分辨率和高分辨率图像计算的形态计量指标总体上仅显示出微弱的依赖性,并且彼此高度相关(R2 = 0.91 - 0.99)。结果表明,全骨和皮质区域的参数非常可重复,而小梁区域的精度略低。然而,如高ICC值所示,所有区域分析方法都非常稳健,证明非常适合应用于近交系,在近交系中由于总体方差小而需要最高的精度。