Lopera Jorge E, Araki Junko U, Kirsch David, Qian Zhong, Brazzini Augusto, Gonzalez Arturo, Castaneda-Zuniga Wilfrido
Department of Radiology, Louisiana State University Health Sciences Center, 1542 Tulane Avenue, New Orleans, Louisiana 70112, USA.
J Vasc Interv Radiol. 2005 Nov;16(11):1539-44. doi: 10.1097/01.RVI.0000177969.33404.16.
With an in vitro model of the inferior vena cava, it was demonstrated that slight tension on the introducer during quick release of the jugular Günther Tulip filter led to less tilting of the filter compared with the technique recommended by the manufacturer. This technique may be useful to minimize significant filter tilting that may be associated with decreased filtration efficiency, and difficulty or impossibility of future filter retrieval.