Hannig C, Dullin C, Hülsmann M, Heidrich G
Department of Operative Dentistry, Preventive Dentistry and Periodontology, University of Göttingen, Germany.
Int Endod J. 2005 Dec;38(12):904-13. doi: 10.1111/j.1365-2591.2005.01033.x.
To detect and to visualize radiographically vertical root fractures in extracted teeth with a prototype of a novel, high resolution, three-dimensional flat panel volume detector computer tomograph (FD-VCT) system.
Five teeth with root fillings and clinical symptoms such as fistulas and isolated periodontal pockets of 8 mm or more were extracted after dental radiography indicating lateral or periapical lesions. Vertical root fractures or cracks were suspected because of the symptoms and clinical findings were evident after extraction in all cases but fracture lines were not visible on routine dental radiographs acquired before extraction. The extracted teeth were explored with a prototype of a FD-VCT. Using the FD-VCT, in all cases vertical root fractures or crack lines could be detected clearly in different views, depiction-modes and cross-sections at a spatial resolution of 140 microm. The evaluation of the fracture lines and teeth could be performed in three-dimensional views. The FD-VCT findings were confirmed by detailed inspection of the extracted teeth.
The FD-VCT is an innovative diagnostic tool for non-destructive, three-dimensional evaluation of extracted teeth in pre-clinical and experimental studies. The FD-VCT allows precise visualization and evaluation of vertical root fractures or cracks in extracted teeth. Clinical application of the system may be possible if technical modifications reduce the exposure dose: the high resolution detector systems of the FD-VCT should be combined with radiation systems that focus the radiation to the area of interest.
使用一种新型高分辨率三维平板容积探测器计算机断层扫描(FD-VCT)系统的原型,检测并通过影像学显示拔除牙齿中的垂直根折情况。
五颗有根管充填且伴有瘘管及8毫米或更深的孤立牙周袋等临床症状的牙齿,在牙科放射成像显示有侧方或根尖周病变后被拔除。由于这些症状,怀疑存在垂直根折或裂纹,且所有病例在拔牙后临床发现都很明显,但拔牙前常规牙科放射片上未见骨折线。使用FD-VCT系统的原型对拔除的牙齿进行检查。使用FD-VCT,在所有病例中,垂直根折或裂纹线在不同视图、描绘模式和横断面上均能以140微米的空间分辨率清晰检测到。骨折线和牙齿的评估可在三维视图中进行。FD-VCT的检查结果通过对拔除牙齿的详细检查得到证实。
FD-VCT是一种创新的诊断工具,可用于临床前和实验研究中对拔除牙齿进行无损三维评估。FD-VCT能够精确显示和评估拔除牙齿中的垂直根折或裂纹。如果通过技术改进降低辐射剂量,该系统可能具有临床应用价值:FD-VCT的高分辨率探测器系统应与将辐射聚焦到感兴趣区域的放射系统相结合。