To find the properties in the oscillatory components of the cutaneous blood flow on the successful free flap, a wavelet transform was applied to the laser Doppler flowmetry (LDF) signals which were measured simultaneously on the surfaces of the free latissimus dorsi myocutaneous flap and on the adjacent intact skin of the healthy limb, of 18 patients. The frequency interval from 0.0095 to 1.6 Hz was examined and was divided into five subintervals (I: 0.0095-0.021 Hz; II: 0.021-0.052 Hz; III: 0.052-0.145 Hz; IV: 0.145-0.6 Hz and V: 0.6-1.6 Hz) corresponding to endothelial metabolic, neurogenic, myogenic, respiratory and cardiac origins. The average amplitude and total power in the frequency range 0.0095-1.6 Hz as well as within subintervals I, II, IV and V were significantly lower for signals measured on the free flap than those obtained in the healthy limb. However in interval III, they were significantly higher. The normalized spectral amplitude and power in the free flap were significantly lower in only two intervals, I and II, yet in interval III they were significantly higher; no statistical significance was observed in intervals IV and V. The distinctive finding made in this study, aside from the decrease of endothelial metabolic processes and sympathetic control, was the significant increase of myogenic activity in the free flap. It is hoped that this work will contribute towards knowledge on blood circulation in free flaps and make the monitoring by LDF more reliable.
为了探寻成功游离皮瓣皮肤血流振荡成分的特性,对18例患者背阔肌肌皮游离皮瓣表面及健康肢体相邻完整皮肤同时测量的激光多普勒血流仪(LDF)信号进行小波变换。研究了0.0095至1.6Hz的频率区间,并将其分为五个子区间(I:0.0095 - 0.021Hz;II:0.021 - 0.052Hz;III:0.052 - 0.145Hz;IV:0.145 - 0.6Hz;V:0.6 - 1.6Hz),分别对应内皮代谢、神经源性、肌源性、呼吸源性和心脏源性。游离皮瓣上测量的信号在0.0095 - 1.6Hz频率范围内以及子区间I、II、IV和V内的平均幅度和总功率显著低于健康肢体上获得的值。然而,在区间III中,它们显著更高。游离皮瓣的归一化频谱幅度和功率仅在区间I和II中显著较低,但在区间III中显著较高;在区间IV和V中未观察到统计学意义。本研究的独特发现,除了内皮代谢过程和交感神经控制的降低外,是游离皮瓣中肌源性活动的显著增加。希望这项工作将有助于了解游离皮瓣中的血液循环,并使LDF监测更加可靠。