射线照相细节与标称焦点尺寸的变化:“焦点效应”

Radiographic detail and variation of the nominal focal spot size: the "focal effect".

作者信息

Katz M C, Nickoloff E L

机构信息

Department of Radiology, Columbia-Presbyterian Medical Center, New York, NY 10032.

出版信息

Radiographics. 1992 Jul;12(4):753-61. doi: 10.1148/radiographics.12.4.1636037.

Abstract

It is not generally appreciated that there is a gradation of the focal spot size from the anode to cathode end of an x-ray field that can dramatically affect radiographic detail from one end of a radiograph to the other. The authors name this gradation the "focal effect." Gradations in radiation intensity and focal spot size were measured from anode end to cathode end of a 14 x 17-inch field and were visually demonstrated with lymphangiograms and line-pair patterns. The degradation in spatial resolution along the anode-cathode axis was found to be as much as 75%, depending on the orientation of the patient with respect to the x-ray tube. Radiographic detail is, therefore, significantly improved (even when the large focal spot is used) by exploiting the focal effect and placing the body part requiring the best radiographic detail at the anode end of the table.

摘要

人们通常没有认识到,在X射线场从阳极到阴极的一端,焦点尺寸存在渐变,这会显著影响射线照片一端到另一端的射线照相细节。作者将这种渐变命名为“焦点效应”。在14×17英寸的视野范围内,从阳极端到阴极端测量了辐射强度和焦点尺寸的渐变,并通过淋巴管造影和线对图案进行了直观展示。发现沿阳极-阴极轴的空间分辨率下降高达75%,这取决于患者相对于X射线管的方位。因此,通过利用焦点效应并将需要最佳射线照相细节的身体部位置于检查台的阳极端,射线照相细节可得到显著改善(即使使用大焦点时也是如此)。