Dysfunction of an ATS valve in the aortic position: the first reported case caused by pannus formation.

作者信息

Teshima Hideki, Aoyagi Shigeaki, Hayashida Nobuhiko, Shojima Takahiro, Takagi Kazuyoshi, Arinaga Kouichi, Yoshikawa Kazuhiro

机构信息

Department of Surgery, Kurume University School of Medicine, 67 Asahi-machi, Kurume, Fukuoka 830-0011, Japan.

出版信息

J Artif Organs. 2005;8(4):270-3. doi: 10.1007/s10047-005-0317-9.

Abstract

A 64-year-old woman underwent aortic valve replacement with a 21-mm Advancing The Standard (ATS) open-pivot mechanical heart valve for bicuspid aortic valve stenosis. In addition to the appearance of a new cardiac murmur, echocardiography performed 3 years after surgery showed a high pressure gradient across the ATS valve and a reduction in the valve orifice area. Cineradiography of the valve revealed restricted leaflet opening. Subsequent multidetector-row computed tomography clearly demonstrated pannus overgrowth on the inflow aspect of the ATS valve. During a repeat operation, subvalvular overgrown pannus was confirmed and the ATS valve was replaced with a bioprosthetic valve. This is the first reported case of prosthetic valve dysfunction resulting from pannus formation in a patient with an ATS valve in the aortic position.

摘要

一名64岁女性因二叶式主动脉瓣狭窄接受了主动脉瓣置换术,植入一枚21毫米的“推进标准”(ATS)开放枢轴机械心脏瓣膜。术后3年进行的超声心动图检查显示,除了出现新的心杂音外,ATS瓣膜上存在高压梯度,瓣膜口面积减小。瓣膜的电影造影显示瓣叶开口受限。随后的多排探测器计算机断层扫描清楚地显示了ATS瓣膜流入面的 pannus过度生长。在再次手术中,证实了瓣下pannus过度生长,并将ATS瓣膜替换为生物瓣膜。这是首例报道的主动脉位置植入ATS瓣膜的患者因pannus形成导致人工瓣膜功能障碍的病例。