Suppr超能文献

主动脉位置的ATS瓣膜瓣叶运动:在19毫米瓣膜中观察到的独特行为。

Leaflet movement of the ATS valve in the aortic position: unique behavior observed in 19-mm valves.

作者信息

Aoyagi Shigeaki, Arinaga Koichi, Fukunaga Shuji, Tayama Eiki, Kosuga Tomokazu, Akashi Hidetoshi

机构信息

Department of Surgery, Kurume University School of Medicine, Kurume, Japan.

出版信息

Ann Thorac Surg. 2006 Sep;82(3):853-7. doi: 10.1016/j.athoracsur.2006.04.088.

Abstract

BACKGROUND

Leaflet opening angles of ATS valves in vivo tend to be less than those reported by the manufacturer, and there is a wide variation in opening angles even among valves of the same size.

METHODS

Leaflet movement of aortic ATS valves was evaluated by cineradiography in 77 patients. The mean period from valve replacement to cineradiography was 40.2 months, and during that period, the ATS valve was replaced due to prosthetic valve obstruction in 1 of the 77 patients.

RESULTS

In 76 patients with a normally functioning ATS valve, the mean opening angle of 19-mm valves (66.3 degrees +/- 2.3 degrees) was significantly less (p < 0.0001) than that of ATS valves 21 mm or larger (72.2 degrees +/- 3.8 degrees in 21-mm, 72.4 degrees +/- 2.8 degrees in 23-mm, and 72.8 degrees +/- 2.9 degrees in 25-mm valves). No significant differences were found in the opening angles of ATS valves 21 mm or larger, and the opening angles exceeded 65 degrees in all 69 patients with valves in this category. In one case of obstruction in a 21-mm valve, the opening angle was 57.5 degrees. The closing angle was 24.9 degrees +/- 1.3 degrees, which corresponded well with the manufacturer's in vitro data.

CONCLUSIONS

The opening angle in the 19-mm ATS aortic valve is significantly less than that in valves 21 mm or larger, and an opening angle of less than 65 degrees probably indicates prosthetic valve obstruction in aortic ATS valves sized 21 mm or larger.

摘要

背景

在体内,ATS瓣膜的瓣叶开放角度往往小于制造商报告的角度,并且即使在相同尺寸的瓣膜之间,开放角度也存在很大差异。

方法

通过电影血管造影术评估了77例患者的主动脉ATS瓣膜的瓣叶运动。从瓣膜置换到电影血管造影术的平均时间为40.2个月,在此期间,77例患者中有1例因人工瓣膜阻塞而更换了ATS瓣膜。

结果

在76例ATS瓣膜功能正常的患者中,19毫米瓣膜的平均开放角度(66.3度±2.3度)明显小于21毫米或更大尺寸的ATS瓣膜(21毫米的为72.2度±3.8度,23毫米的为72.4度±2.8度,25毫米的为72.8度±2.9度)(p<0.0001)。21毫米或更大尺寸的ATS瓣膜的开放角度没有显著差异,在所有69例此类瓣膜患者中,开放角度均超过65度。在1例21毫米瓣膜阻塞的病例中,开放角度为57.5度。关闭角度为24.9度±1.3度,与制造商的体外数据非常吻合。

结论

19毫米的主动脉ATS瓣膜的开放角度明显小于21毫米或更大尺寸的瓣膜,开放角度小于65度可能表明21毫米或更大尺寸的主动脉ATS瓣膜存在人工瓣膜阻塞。