Boden Catherine, Sit Arthur, Weinreb Robert N
Hamilton Glaucoma Center, Department of Ophthalmology, University of California at San Diego, La Jolla, USA.
J Glaucoma. 2006 Feb;15(1):30-4. doi: 10.1097/01.ijg.0000196654.77836.61.
To evaluate the accuracy of a prototype electronic device for recording eye drop usage.
Ten volunteers were randomly assigned to one of five usage patterns designed to mimic common patterns of use in glaucoma patients from 100% compliant to 50% compliant. All participants agreed to adhere to a pre-determined "dosing" schedule for 15 days using the monitoring/reminder device to instill artificial tears. Participants also recorded drop usage in a diary. The main outcome measures were device accuracy and reproducibility. Device accuracy was defined as the magnitude of the difference between the diary and device output for three variables: date, number of drops, and instillation time.
Date stamping by the device was 100% accurate. The mean +/- SD time difference between the device and the diary was -2.0 +/- 19.7 minutes when data from all participants was pooled. In seven of the ten participants, the device did not record at least one drop. The mean +/- SD difference in the number of drops recorded by the device minus the diary was 0.16 +/- 0.97 when data from all participants was pooled.
The prototype compliance reminder/monitoring device may underestimate compliance in some patients. The date and time stamping mechanisms were generally accurate and reproducible.
评估一种用于记录眼药水使用情况的原型电子设备的准确性。
10名志愿者被随机分配到五种使用模式中的一种,这些模式旨在模拟青光眼患者从100%依从到50%依从的常见使用模式。所有参与者同意使用监测/提醒设备滴入人工泪液,按照预先确定的“给药”时间表持续15天。参与者还在日记中记录眼药水的使用情况。主要观察指标为设备的准确性和可重复性。设备准确性定义为日记与设备输出在三个变量上的差异幅度:日期、滴数和滴入时间。
设备的日期标记准确率为100%。汇总所有参与者的数据后,设备与日记之间的平均±标准差时间差为-2.0±19.7分钟。在10名参与者中的7名中,设备未记录至少一滴眼药水。汇总所有参与者的数据后,设备记录的滴数减去日记记录的滴数的平均±标准差差异为0.16±0.97。
该原型依从性提醒/监测设备可能会低估某些患者的依从性。日期和时间标记机制总体上准确且可重复。