Lu Bing, Jonsson Peter, Blomberg Sverker
Analytical Chemistry, Process R&D, AstraZenca R&D, Södertälje, Sweden.
J Chromatogr A. 2006 Jun 30;1119(1-2):270-6. doi: 10.1016/j.chroma.2005.11.110. Epub 2005 Dec 27.
A reversed phase ion-pair gradient liquid chromatographic method has been developed and validated for purity determination of the hydrophilic compound 2,4-disulfonic acid benzaldehyde di-sodium salt (2,4-DSAD) containing both hydrophilic and more lipophilic related impurities. Mixtures of acetonitrile-phosphate buffer containing tetrahexylammonium hydrogen sulfate as the ion-pairing reagent were used as the mobile phase. A linear gradient, which generated simultaneous change in the concentration of organic modifier, buffer concentration and the concentration of ion-pairing reagent, was applied. The method allows detection of impurities at low levels (0.01% w/w). Excellent repeatability for both retention time (RSD< or =0.3%, n = 6) and detector response (RSD = 0.03%, n = 6 for the main peak and RSD = 6%, n = 6, for an impurity at 0.01 area% level) was obtained. The method was shown to be robust for routine analysis and has been successfully transferred to the quality control laboratory.
已开发并验证了一种反相离子对梯度液相色谱法,用于测定含有亲水性和更多亲脂性相关杂质的亲水性化合物2,4-二磺酸苯甲醛二钠盐(2,4-DSAD)的纯度。以含有硫酸四己铵作为离子对试剂的乙腈-磷酸盐缓冲液混合物作为流动相。采用线性梯度,使有机改性剂浓度、缓冲液浓度和离子对试剂浓度同时变化。该方法能够检测低水平杂质(0.01% w/w)。保留时间(RSD≤0.3%,n = 6)和检测器响应均具有出色的重复性(主峰RSD = 0.03%,n = 6;0.01面积%水平的杂质RSD = 6%,n = 6)。该方法在常规分析中表现出稳健性,并已成功转移至质量控制实验室。