赞比亚HIV感染者中,使用Dynal T4 Quant CD4检测法进行延迟手动CD4检测时,Cyto-Chex血液防腐剂的效果。

Efficacy of Cyto-Chex blood preservative for delayed manual CD4 testing using Dynal T4 Quant CD4 test among HIV-infected persons in Zambia.

作者信息

Truett April A, Letizia Andrew, Malyangu Evans, Sinyangwe Frank, Morales Brandi N, Crum Nancy F, Crowe Suzanne M

机构信息

Division of Infectious Diseases, Naval Medical Center, San Diego, CA 92134, USA.

出版信息

J Acquir Immune Defic Syndr. 2006 Feb 1;41(2):168-74. doi: 10.1097/01.qai.0000191282.77954.fc.

Abstract

BACKGROUND

Manual CD4 tests such as Dynal T4 Quant (Dynabeads, Dynal Biotech, Oslo, Norway) are less expensive alternatives to flow cytometry in resource-limited countries. Whereas blood preservatives have proven useful for stabilizing blood samples to allow delayed CD4 testing by flow cytometry, they have not been verified for manual tests. A method for preservation of blood prior to manual CD4 testing is needed for long-distance transport or sample batching.

METHODS

Blood from HIV-positive Zambian military beneficiaries was mixed (1:1) with Cyto-Chex (Streck Laboratories, La Vista, NE) blood preservative, and the blood was stored at refrigerated, ambient, and incubator (37 degrees C) temperatures prior to Dynabeads CD4 testing at 0, 3, 6, and 9 days after collection. Baseline flow cytometry and Dynabeads testing without preservative were performed for comparison.

RESULTS

Twenty-seven patient samples were analyzed. Dynabeads vs. flow cytometry had a correlation coefficient (r) of 0.84. There was excellent correlation (r = 0.96) between baseline Dynabeads testing and Cyto-Chex-preserved samples. Refrigerated samples showed strong correlation with baseline Dynabeads (r = 0.93-0.95) on days 3, 6, and 9 without decline in CD4 count (P = 0.73). Samples stored at ambient temperature yielded inferior results (r = 0.76-0.81), with a significant decline in CD4 count by day 3 (P < 0.001). The incubator arm had especially poor correlation (r = 0.30-0.49).

CONCLUSIONS

Addition of Cyto-Chex to peripheral blood (1:1) adequately preserves refrigerated blood samples for up to 9 days for subsequent testing with Dynabeads CD4 test. Cyto-Chex, however, cannot be recommended for delayed Dynabeads CD4 testing with storage at 37 degrees C or ambient temperatures in tropical areas similar to the site of this study.

摘要

背景

在资源有限的国家,诸如Dynal T4 Quant(挪威奥斯陆Dynal生物技术公司的磁珠)等手动CD4检测方法是比流式细胞术更经济的替代方法。虽然血液防腐剂已被证明可有效稳定血液样本,以便通过流式细胞术进行延迟CD4检测,但尚未在手动检测中得到验证。对于长途运输或样本批量处理,需要一种在手动CD4检测前保存血液的方法。

方法

将来自赞比亚HIV阳性军事受益人的血液与Cyto-Chex(美国内布拉斯加州拉维斯塔市的Streck实验室)血液防腐剂按1:1混合,并在采集后第0、3、6和9天进行Dynal磁珠CD4检测前,将血液分别储存在冷藏、室温及培养箱(37摄氏度)温度下。同时进行无防腐剂的基线流式细胞术和磁珠检测以作比较。

结果

分析了27份患者样本。磁珠法与流式细胞术的相关系数(r)为0.84。基线磁珠检测与经Cyto-Chex保存的样本之间具有良好的相关性(r = 0.96)。冷藏样本在第3、6和9天与基线磁珠检测显示出强相关性(r = 0.93 - 0.95),CD4计数无下降(P = 0.73)。室温储存的样本结果较差(r = 0.76 - 0.81),到第3天CD4计数显著下降(P < 0.001)。培养箱组的相关性尤其差(r = 0.30 - 0.49)。

结论

向外周血中添加Cyto-Chex(1:1)可充分保存冷藏血液样本长达9天,以便随后用Dynal磁珠CD4检测。然而,对于类似于本研究地点的热带地区,在37摄氏度或室温下储存进行延迟的Dynal磁珠CD4检测,不推荐使用Cyto-Chex。